PIP can be set but the tidal volume cannot. A full list of how each
setting is treated in each mode can be found below:
Additional Settings
O
2
Conserve
is used when an oxygen concentrator is employed as the
oxygen source, or when the user wishes to conserve the remaining
oxygen in a depleting cylinder. This function limits the amount of
oxygen used during inspiration (either through the hose connector or
the barbed tree connector) to 10 LPM. See USING THE CCV WITH A
CONCENTRATOR OR FLOWMETER
for more information.
Manual Breath
delivers a single breath at the current ventilator settings
in volume and pressure (but not in CPAP) modes.
Control Lock
freezes the buttons on the control panel. When activated,
a blue light will illuminate beneath the button. The Alarm Silence/Reset
button is still active.
Control Lock will automatically come on if Apnea
Ventilation Mode is engaged. To change ventilation modes or settings,
you must first disable the lock.
Parameter/mode Adjustment
arrows are used to adjust settings or
choose between programmed Custom Modes. For more information on
changing settings and modes see
SELECTING SETTINGS AND MODES
.
Ventilation Mode
Volume Modes Pressure Modes CPAP
I. Time
Required
Required
N/A
BPM
Required
Required
N/A
%O
2
Required
Required
Required
PEEP/CPAP
Optional
Optional
Required
PIP
N/A
Required
N/A
Vt
Required
N/A
N/A
Pressure Support
Optional
Optional
Optional
Trigger Sensitivity
Required
Required
Required
High Pressure Alarm
Required
Required
Required
Low Pressure Alarm
Required
Required
Required
O
2
Conserve
Optional
Optional
Optional