background image

10

8

Err  11

Load-cell problem

9

Err  13

Resolving capability is too high

10

Err  14

Damaged inside indicator

11

Loch

Overload and locked

RECHARGEABLE LITHIUM BATTERY

Battery inside of scale is Lithium battery. Please charge it after running 
down for 7 hours the first three times. Please charge the battery for timing 
if not use it often. 

The battery is an easily exhausted product. And it is not granted free 
guarantee. 

MAINTENANCE

1.  To guarantee indicator clarity and using life, the indicator shouldn’t be 

placed directly under sunshine and should be set in the plain space.

2.  The indicator can’t be placed into the place where the dust pollution and 

vibration are serious.

3.  Load cell should connect with indicator reliably, and the system should 

be well connected into ground. The indicator must be protected from 
high electrical field and high magnetic field.
In order to protect the operator, indicator and relevant device, you should 
mount lightning rod in thunderstorm frequently happening area.

•  Load cell and indicator are static sensitive device, you must adopt 

anti-static measures.

4.  It is strictly forbidden to clean the case of indicator with intensive solvents 

(for example: benzene and nitro oils).

Summary of Contents for C6-10T

Page 1: ...SERIES C6 3T 5T 10T 15T 20T OPERATION MANUAL ...

Page 2: ... Front and back view of the indicator 4 Keys function 4 Operation 5 Turning on 5 Zero function 5 Tare 5 Manual accumulation 5 Check accumulation 6 Clear accumulation 6 Parameters setup 6 Calibration 8 Error indication 9 Rechargeable lithium battery 10 Maintenance 10 Guarantee 12 ...

Page 3: ... times second Display Display 0 999999 5 digits LCD LED 7 indicators Division Value 1 2 5 10 20 50 optional Ambient Environment Distance between scale and indicator 150m Power Supply DC 3 7V Ambient Temperature 0 40 ºC Storing Temperature 25 ºC 55 ºC Ambient Moisture 85 RH Warm up Time 10 15 分钟 ...

Page 4: ... than 5s it enters setup mode All the setting in this mode should be recorded with the calibration header on or the data will lose Tare function in weighing status Zero function in weighing status For selecting Weighing accumulation Long press for 1sec to turn on Long Press for 3sec to turn off ...

Page 5: ... open both indicator and scale change the number of transcoder inside with same number and restart Tare At weighing mode press to deduct displaying weight while it is positive and stabilization light is on At no load status of the scale press to deduct tare and then tare light is off Manual accumulation At weighing mode press it displays ADD 01 01 means times of accumulation the Max times is 99 an...

Page 6: ...he current accumulation PARAMETERS SETUP In weighing status hold for more than 5s and entering setup mode P mode all the setting in this mode should be recorded with the calibration header on or the data will lose P1 P13 parameters are available press to switch and press to select parameter Detail for parameter 1 P1 x kg Lb shift X 1 kg mode default X 2 Lb mode 2 P2 x Not available for this model ...

Page 7: ...e X 1 No power saving X 2 Power saving available X 3 None 7 P7 x Zero tracking Scope X 1 0 5e X 2 1 0e X 3 1 5e default X 4 2 0e X 5 2 5e X 6 3 0e X 7 3 5e 8 P8 x Zero key Scope X 1 2 FS X 2 4 FS X 3 10 FS default X 4 20 FS 9 P9 x Zero scope upon start X 1 2 FS X 2 4 FS X 3 10 FS default X 4 20 FS ...

Page 8: ...lue Press to select division value then press to confirm CALIBRATION Turn on the crane scale first then hold key and power on the indicator secondly it enters calibration mode Step Operation Display Explanation 1 Press to choose division value several times d Division value 1 2 5 10 20 50 optional press to confirm e g 2 2 Press to choose decimal point position several times P Decimal point optiona...

Page 9: ...weights loaded Press to choose the digit bit press to select the digit Until the input value is same as the loaded weight and the digit bit is at the highest bit press to confirm when the stable light is on 7 Calibration finished ERROR INDICATION 1 Err 1 AD value is too small when calibration 2 Err 2 Zero point is out of range when calibration 3 Err 3 Zero point is out of range when power on 4 Err...

Page 10: ...t be placed directly under sunshine and should be set in the plain space 2 The indicator can t be placed into the place where the dust pollution and vibration are serious 3 Load cell should connect with indicator reliably and the system should be well connected into ground The indicator must be protected from high electrical field and high magnetic field In order to protect the operator indicator ...

Page 11: ...ction may happen 8 The storage is not granted the free repair guarantee because it is easily exhausted products In order to prolong using life please charge the cell fully before using it If you don t use the indicator for a long time you must charge the cell every two month and for eight hours each charging time Moving or installation must be carefully taken and must avoid strong vibration impact...

Page 12: ...ation defect of the material During this period GRAM PRECISION SL covers the manpower and the spare parts necessary to repair the scale This guarantee does not cover the failures caused by an inappropriate use or overload The guarantee does not cover the freight cost transport necessary to repair the balance ...

Page 13: ...13 NOTES ...

Page 14: ...14 NOTES ...

Page 15: ...15 NOTES ...

Page 16: ...Gram Precision S L Travesía Industrial 11 08907 Hospitalet de Llobregat Barcelona Spain Tel 34 902 208 000 34 93 300 33 32 Fax 34 93 300 66 98 comercial gram es www gram group com 003 19072018 ...

Reviews: