9
3
Setup F.S.
[FULL ]
Input F.S.:Press
to choose
the digit bit; press
to select
the digit; press
to confirm
the input F.S. e.g.: 30000
4
Zero point calibration:
Press
when the stable
light is on
[noLoAd] Assure there’s no loading
5
Load standard weights.
6
F.S. calibration: (Press
directly if the value to be
input is same as F.S. when
stable light is on)
[AdLoAd]
Input weight of standard weights
loaded: Press
to choose
the digit bit; press
to select
the digit. Until the input value is
same as the loaded weight and
the digit bit is at the highest bit,
press
to confirm when the
stable light is on.
7
Calibration finished.
ERROR INDICATION
1
Err 1
AD value is too small when calibration.
2
Err 2
Zero point is out of range when calibration.
3
Err 3
Zero point is out of range when power on
4
Err 5
Weight input is 0 when full scale calibration
5
bAt-Lo
Indicator out of battery, please charge!
6
Err 8
Mistakes in inputting F.S.
7
Err 10
Resolving capability is too high