Platform Shaker CPS-20
Page 8
Operating manual
5. Operation
Figure 1. Control panel of CPS-20
5.1.
Connect the external power supply to the mains. Display turns on.
5.2.
Place samples on the unit platform.
5.3.
Setting the parameters
. Press the
Select
key (fig. 1/1) to choose the parameter to change. Each
pressing of the
Select
key consecutively activates the parameters. The active parameter is flashing.
Use the ▼ and ▲ keys (fig. 1/2) to set the necessary value. Pressing the key for more than 2 s
increases the speed of value change.
5.3.1. Set the required working time interval in hours and minutes, the increment is 1 minute.
5.3.2. Set the required shaking speed, the increment is 10 rpm.
5.4.
Press the
Run Stop
key (fig. 1/3). The platform starts rotation, indication RUN appears on display
and the timer in the lower line of the display starts counting the time interval.
5.5.
After the timer reaches the set time, the platform motion will stop and the flashing indication STOP,
accompanied by the repetitive sound signal, will appear in the lower line of the display. Press the
Run
Stop
key to shut down the signal.
5.6.
The unit can be stopped before the set time elapses if necessary by pressing the
Run Stop
key.
Press the
Run Stop
key to repeat the operation with the same working time and speed.
5.7.
If the working time is not set (or is reset) and the Time indicator on display shows OFF, pressing the
Run Stop
key will start continuous operation of the unit until the
Run Stop
key is pressed.
5.8.
The platform motion can be stopped at any time by pressing the
Run Stop
key. In this case, the
program realization and the platform motion will stop and the unit will switch into the STOP mode.
5.9.
After finishing the operation disconnect the external power supply from the mains.