Operating instructions

High–Speed 

Multi Plate Shaker

MPS-1

For version 

V.1G

W

Summary of Contents for MPS-1

Page 1: ...Operating instructions High Speed Multi Plate Shaker MPS 1 For version V 1GW ...

Page 2: ...tions Version 1 01 November 2013 1 Safety 3 2 General Information 4 3 Getting Started 5 4 Operation 7 5 Specifications 9 6 How to choose the right mode for mixing 11 7 Maintenance 12 8 Guarantee and service 13 9 Declaration of Conformity 14 ...

Page 3: ...upply connector are easily accessible during use Disconnect the unit from the electric circuit before moving Turn off the unit by disconnecting the external power supply from the power socket If liquid penetrates into the unit disconnect it from the external power supply unit and have it checked by a repair and maintenance technician Do not operate the unit in premises where condensation can form ...

Page 4: ...d Multi Plate Shaker provides Pulse mode featuring a series of 3 second pulses each increasing linearly in rpm until the set speed is reached Activated by a separate key on the front panel this motion is repeated until the timer stops This mode provides a constant state of resuspension of particles as the acceleration is constantly changing Universal platform holder accommodating microplates PCR p...

Page 5: ...te shaker with universal platform plate holder u 1 piece external power supply unit 1 piece Operating Instructions Certificate 1 copy Optional accessories platform P 2 24 for 24 tubes 1 5 2 ml v on request platform P 05 32 for 32 tubes 0 5 ml w on request platform P 02 05 for 24 tubes 0 5 ml and 48 tubes 0 2 ml x on request platform P 02 96 for 96 tubes 0 2 ml or semi unskirted PCR plate y on requ...

Page 6: ...upply unit into the 12 V socket at the rear side of the unit 3 4 Platform or plate installation Install the platform or plate by inserting it in the universal platform plate holder fig 1 Fig 1 Platform or plate installation FRONT Page 6 MPS 1 Operating instructions Version 1 01 November 2013 ...

Page 7: ...ximum fill volume for efficient mixing 4 1 Connect the external power supply unit to electric circuit and turn ON the power switch located on the rear panel of the unit position I 4 2 Place the plate or optional platform on the universal platform plate holder and push it inside firmly fig 1 Insert the tubes into the sockets of the optional platform Caution Before mixing it is recommended to use CU...

Page 8: ... to turn off the signal 4 8 The unit can be stopped before the set time elapses if necessary by pressing the Start Stop key The Time indicator will show the set time interval Press the Start Stop key to repeat the operation with the same working time and speed 4 9 If the working time is set to zero and the display shows 00 00 pressing the Start Stop key will start continuous operation of the unit ...

Page 9: ...level max 65 dBA Maximum continuous operation time 8 hrs Dimensions 225x215x150 mm Input current power consumption 12 V 800 mA 10 W External power supply unit input AC 100 240 V 50 60 Hz output DC 12 V Weight 5 1 kg Accurate within 10 n in cold rooms incubators and closed laboratory rooms at ambient temperature from 4 C to 40 C in a non condensing atmosphere and maximum relative humidity 80 for te...

Page 10: ...l platforms platform for 24 tubes 1 5 2 ml P 2 24 platform for 32 tubes 0 5 ml P 05 32 platform for 96 tubes 0 2 ml or semi unskirted PCR plate P 02 96 platform for 24 tubes 0 5 ml and 48 tubes 0 2 ml P 02 05 Grant is committed to a continuous programme of improvement and reserves the right to alter design and specifications of the equipment without additional notice Page 10 MPS 1 Operating instru...

Page 11: ...atforms Preset mode 1000 rpm 1800 rpm 2600 rpm Deep well plate full Height 44 mm 2000 µl Deep well plate full Height 29 mm 500 µl Immunoplates full Height 15 mm 300 µl PCR plate skirted full Height 15 mm 200 µl Platform Р 2 24 24 x 2 ml 1 5 ml Platform Р 05 32 32 x 0 5 ml full Platform Р 02 05 24 x 0 5 and 48 x 0 2 ml full Platform P 02 96 strips PCR plate 96 x 0 2 ml full SOFT MEDIUM HARD Tubes m...

Page 12: ...solvents Standard ethanol 75 or other cleaning agents recommended for cleaning of laboratory equipment can be used for cleaning and decontamination of the unit Before using any other cleaning decontamination method other than that recommended check with our service department or distributor if outside of the UK that the proposed method will not damage the equipment all Grant laboratory products ar...

Page 13: ...ons this product is guaranteed for TWO YEARS against faulty materials or workmanship 8 2 Service For service return for repair to our Service Department in the UK or in other countries to our distributor 8 Guarantee and Service Page 13 MPS 1 Operating instructions Version 1 01 November 2013 ...