Thermo shaker 

PHMP/ PHMP-4

For version V.2GW

Summary of Contents for PHMP

Page 1: ...Operating instructions Thermo shaker PHMP PHMP 4 For version V 2GW ...

Page 2: ...hermo shaker PHMP PHMP 4 Operating instructions Version 2 03 July 2013 Contents 1 Safety 3 2 General Information 5 3 Getting Started 6 4 Operation of PHMP PHMP 4 7 5 Specifications 9 6 Guarantee and service 10 ...

Page 3: ...cuit Use only cleaning and decontamination methods recommended by the manufacturer Do not make modifications to the design of the unit ELECTRICAL SAFETY Connect only to the external power supply unit with voltage corresponding to that on the serial number label Use only the external power supply unit provided with this product Ensure that the switch and external power supply unit are easily access...

Page 4: ...ith aggressive or explosive chemical mixtures Do not operate the unit if it is faulty or has been installed incorrectly Do not use outside laboratory rooms Do not check the temperature by touch Use a thermometer BIOLOGICAL SAFETY It is the user s responsibility to carry out appropriate decontamination if hazardous material is spilt on or penetrates into the equipment Version 2 03 July 2013 ...

Page 5: ...of microplates v Microplate Shaker for operation in the cold room or other conditions which do not require temperature stabilisation w Thermo Shaker for immunochemistry and molecular diagnostics where stringent requirements of repeatability and precision are necessary An external 12 V power supply unit is used to power the unit This makes it safe for use in the cold room where condensation may cau...

Page 6: ... rubber drive belt 2 pieces External power supply unit 1 piece Operating instructions Declaration of Conformity 1 copy 3 3 Set up Place the unit onto an even horizontal non flammable surface away from any flammable materials not less than 30 cm Plug the external power supply unit into the socket at the rear side of the PHMP and position the unit so that there is easy access to the power switch and...

Page 7: ...in hours and minutes increment 1 min Pressing the key for more than 3 sec will increase the increment Setting speed RPM 4 4 Using the p and q keys Fig 1 v set the required speed increment 10 RPM Pressing the key for more than 3 sec will increase the increment Setting temperature T ºC 4 5 Using the p and q keys Fig 1 w set the necessary temperature increment 0 1ºC Pressing the key for more than 3 s...

Page 8: ...e Thermo shaker timer indicator will start counting up the time interval in the lower line Actual until the RPM RUN STOP key is pressed again Please note The platform temperature will be constantly maintained in accordance with the set temperature This allows the devise to be used again without pre heating 4 10 The timer can be reset during operation if required Press the TIME RUN STOP key once Fi...

Page 9: ...y 16x2 signs LCD Digital time setting range 1 min 96 hrs non stop Max continuous operation time 96 hrs recommended interval between operation sessions not less than 8 hrs Time setting resolution 1 min Max height of miocroplate 18 mm External power supply input AC 100 240 V 50 60Hz output DC 12V PHMP PHMP 4 Current power consumption 12 V DC 3 3 A 40 W 12 V DC 4 15 A 50 W Platform loading capacity 2...

Page 10: ...5 years or 2000 hours of operation time 1 Disconnect the external power supply unit from the device 2 Remove 4 fixation screws on the device base and remove the bottom plate 3 Replace the rubber drive belt fig 2 4 Re assemble the device 6 2 2 Spare parts Rubber drive belt 117x5x0 6 mm External power supply unit part no 18020 input AC 100 240 V 50 60 Hz output DC 12 V 6 3 Cleaning disinfection Stan...