WARNING
Setting the liquid level sensor to OFF means that the Grant digital ultrasonic bath will
operate as if the bath were full of liquid.
This means that without water, the heaters will continue to operate and the surface
inside the bath will become extremely hot. It is therefore the operator’s responsibility
to ensure that the tank is filled with sufficient water when the liquid level sensor is
disabled.
If the Grant digital ultrasonic bath operates with insufficient liquid, the heaters will be
disabled as a safety feature and the following screens will be displayed:
To continue using the Grant digital ultrasonic bath, refill with liquid up to the liquid level
sensor and switch the unit off and on again.
If the bath is run dry this will invalidate the warranty on the Grant digital
ultrasonic bath.
A warning label is included in the literature pack that may be applied to the
unit if deemed necessary following a risk assessment
(see ISO 13732-1:2006).