When the Grant digital ultrasonic bath power is turned on and the liquid level is too
low, the following screen will display.
The level sensor can be disabled for specialist cleaning applications where deionised
(DI) water is used (see page 19).
When the bath is filled with liquid to the correct level, you will be given the option to
degas (see page 25).
By choosing YES, your Grant digital ultrasonic bath will run an automatic degas cycle.
The following screen will display.
The degassing process is to remove all bubbles from the water as these impede the
cleaning performance. Once all of the bubbles have come to the surface, complete the
degassing cycle by pressing the key.
Note: If the machine detects low liquid level for more than 5 seconds, it is assumed
that the bath has been filled with fresh water, and you will be given the option of
running a degas cycle.