CRF-1 Operating Manual
Page 18 of 60
33801-V4 DMN S90
6.3. Running a profile
Profiles can be set up and run using the
Monitor
page.
Figure 17: Software: Choosing Profile
6.3.1. Selecting the profile & details
Before running a cooling profile check the following are correct:
•
The cooling profile
•
The logging interval
•
The name and storage folder of the log file
6.3.1.1. Choosing the profile to run
In the
Profile Details
window choose a profile from those loaded on the CRF-1 from the
Profile
options (e.g. Cryo1 in Figure 17). The specific temperature details of the chosen
profile can be viewed by clicking on the
View Details
box. This will help ensure you have
chosen the correct profile before it is run.
6.3.1.2. Setting the log file parameters
Set the
Log Interval
,
Log Directory
and
Log Filename
in the
Profile Details
window. The
log interval is normally set at 2 seconds for typical cryopreservation applications with run
lengths up to approximately three hours (e.g. stem cell, blastocysts, mouse embryo, somatic
cells etc).
Note: Shorter logging intervals and long cryopreservation profiles result in very large log files
that may not provide any more information. For applications that need prolonged freezing
stages, the interval can be increased to up to 60 minutes between logging points dependent
on application.