CRF-1 Operating Manual
Page 16 of 60
33801-V4 DMN S90
6.2. Establishing communications with the freezer unit
Ensure the computer is connected to the unit using the serial communications lead or USB to
Serial converter.
On the Monitor panel, click
on the ‘Connect’ button as shown below. The ‘Comms Status’ at
the bottom of the panel will
change to ‘Status: Connected’. Other details such as the unit
model and serial number will also be displayed. If the software fails to connect to the unit, see
Appendix A1.3 for help.
Figure 15: Software: Connect
The freezer unit will remain connected until the Disconnect button is pressed or the unit is
switched off.
The system will be ready to use when the freezer unit plate temperature has stabilised at the
Start Temperature. This may take approximately 5 minutes if the unit has only just been
switched on
When the freezer has reached a stable temperat
ure the Cryocooler Status will display ‘Status:
R
eady’. See figure below. The ‘Ready’ green light will also illuminate on the freezer front
panel.