CRF-1 Operating Manual
Page 48 of 60
33801-V4 DMN S90
8.7. Cryovial Cryopreservation
Profile name: Cryo1
Applicable models: H02, H05, H06
Start temperature
20°C
Cool
from 20°C to -5°C at 2°C min
-1
Cool
from -5°C to -7°C at 1°C min
-1
Hold for nucleation
at -7°C for 10 minutes
Cool
from -7°C to -80°C at 1°C min
-1
Hold
at -80°C
Transfer to storage
Figure 53: Pre-programmed cryovial freezing cycle
8.8. Test and Calibration profiles
Profile name: TCal, TP1, TP2
Applicable models: H00, H01, H03, H07, H08
Profile name: TP3, TP4
Applicable models: H02, H05, H06
These profiles are for testing and calibrating the freezer unit. Please do not edit or remove
these profiles. For further information on how to test and calibrate your freezer unit contact
Grant Instruments.