Version 12
28123
Page 15
www.grantinstruments.com
Please Note:
The heatsink material is developed to absorb large amounts of heat energy, as a result of this it has
a relatively low melting point.
So it is essential the heatsink should not be used if the heatsink material is in liquid form.
The material will change phase from a solid to a liquid when the temperature of the heatsink is above
32°C.
Therefore the heatsinks are required to be kept cool to ensure they offer the maximum protection
before use. Ideally they should be at 20 -
22°C. In warm ambient temperatures i.e. above 25°C then
it is recommended that the heatsink be cooled/chilled prior to use. A possible solution is a
refrigerator at 7°C) . Failure to cool the barrier between use will result in the logger being permanent-
ly damaged.
Re-
crystallisation (change from liquid to solid) will start between 24 and 26°C; cooling in a freezer or
ice bath / cool water bath will speed up this process.
6.3 TBHS-1 Heatsink User Information
Hazardous Information:
Hazard identification of heat sink material: none
First aid measures:
General information:
No particular measures are required
After skin contact:
Wash immediately with plenty of water
After eye contact:
Rinse immediately under running water for several minutes with eyelids held open
and seek medical advice
If swallowed:
Seek medical advice
Summary of Contents for OQ610-S
Page 1: ...ES IT DE FR EN ...
Page 19: ...28123 Version 12 Page 18 www grantinstruments com EN Personal Notes ...
Page 37: ...28123 Version 12 Page 18 www grantinstruments com FR Notes persannelles ...
Page 55: ...28123 Version 12 Seite 18 www grantinstruments com DE Notizen ...
Page 73: ...28123 Versione 12 Pagina 18 www grantinstruments com IT Note personali ...
Page 91: ...Version 12 28123 Page 18 www grantinstruments com Notes ...