33784 V1.0
Yoyolog User Guide
www.grantinstruments.com
Page 32
9
Troubleshooting
Typical error messages that may occur during operation are listed below.
Try to solve the problem with the following help list before contacting your distributor.
A) No logger is displayed
First, check if the logger cable is correctly connected to the USB Port and
to the logger. Secondly, uninstall the software and re-install the complete
package (software/USB-driver). Administration rights are required dur-
ing the installation process. If the problem still exists there may be a fault
with the logger.
B) Unknown Logger is displayed
The software being used may be an old version, which doesn’t recog-
nize the new hardware. Check with the local dealer for any upgrades.
C) Cannot contact logger
If the error message continues to show after checking that the loggers is
connected correctly and appears in the YoyoView software. This could
be due to the logger having a flat or missing battery. Replace the battery
and verify initial function test as follows:
After having inserted the battery the LED on logger front should flash in
all three colours (LED chaser). This shows the logger is operating. Now,
try contacting the logger. If problem still exists there may be a problem
with the logger hardware, contact the local dealer.
D)
Logger
stops
taking
readings
after
a
period
logging.
Please check the logger setup for any settings selected for a stop
date/time, change the setup if necessary. Otherwise this could be due to
a flat battery. Check battery voltage in YoyoView (see chapter 4.1) and
possibly replace battery.
E)
Humidity/Temperature readings continuously show -40°C/+120°C .
Humidity/Temperature-Sensor is faulty or the cable has been damaged.
F)
Other sensors continuously show their Max/Min:
Sensor or sensor cable could be damaged.