GL820-UM-251-9370
5-12
5. INSPECTION AND CHECK PROCEDURES
(4) Press the [CH GROUP] key to display from the CH11 to the CH20.
(5) Press the [START/STOP] key to measure.
(6) Wait until the "Finished" message is displayed, and then display the values as shown below.
Message
Confirm that this value is within
the below ratings after data has
been captured.
(7) Check that the level of measurement is within below rating.
RTD precision ratings
Range
Input resistor
Reference
Temperature
Factory Rating
Product Rating
Pt100
350 Ω
715.3 ˚C
+714.5 ˚C to +716.1 ˚C
+714.3 ˚C to +716.3 ˚C
Pt1000
2500 Ω
408.8 ˚C
+407.8 ˚C to +409.2 ˚C
+408.0 ˚C to +409.6 ˚C
Summary of Contents for GL820
Page 1: ...midi LOGGER SERVICE MANUAL 820 GL820 UM 251 GL820 UM 251 02 9370 ...
Page 2: ......
Page 8: ...GL820 UM 251 9370 1 2 1 CONFIGURATION ...
Page 12: ......
Page 36: ......
Page 56: ......