CPT20 USER MANUAL
10
T 7 4 2 1 1
means that the scale has the capacity of 15 lb with dual range. The price is
listed as 0.00, G value is available for standard calibration, and P2 starts from P11.
Weight Calibration
The SW1 calibration switch on the main board switched to ON position
Press the T key and the 2 key in weighing mode to start Weight Calibration
Weight Calibration will not be accessible when SW1 calibration is on CAL_OFF
position
The scale reads the calibration zero for 1 second, do not place anything onto the scale
Place the weight mass or use 0 ~ 9 < > keys to change the weight
value. Place the weight mass and place the ENTER key
Note: For max capacity of 15 lb, you can use 10 lb weight to do the
calibration. It is suggested to use at least 1/3 of maximum up to max for the
calibration purpose.
The scale returns to weighting mode.
To calibrate only the initial zero and not weight calibration, just press the enter key
G Value Calibration and how to view G value :
Press the T key and he 3 key in weighing mode to start G value calibration
or to view G value
Notes for G values:
1. 10 G values can be entered
2. Only one G value can be entered when the calibration switch is at
CAL_ON position
3. Make sure the G value entered is correct to avoid inaccurate weighing
4. The G value will be denied when the value is greater than 9.83217 (Polar
0 1 5 . 0 0 0
E S C
0 . 0 0 0
0
0 . 0