L-SOLUTION 100 Ex Fibre machine
. 82
First contact
Autofocus adjustment
In order to adjust the autofocus, the machine must imperatively be equipped with a lens
of 3" and you must use only marked measuring wedge 3".
Do not use the controls or settings and do not carry out procedures other than those
specified in the use manual, to avoid risking exposure to dangerous radiation.
After an incident or a bad handling with the head of the LASER machine, the autofocus can go wrong .
To solve this problem, a simple procedure can be effected by the user :
Remove all the tools or the objects being on the table because the head has to move in
maximal speed towards the center of the table to make this adjustment.
1. Switch off the machine (main stop button (L) in 0 position)
2. Close the door
3. Switch on the machine (main stop button (L) in I position)
The following messages display on the L.C.D. screen :
SET UP
FPGA
X.X
INIT LASER FIBRE ...
SET UP
The machine will then issue a "beep" and the following screen will remain displayed for 3 seconds:
<< GRAVOGRAPH >>
FPGA
X.XX
LCD
SOFT
X.XX
32 Mo
BOOT X.XX
4. While this message is displayed, press
SET LASER TYPE
TICKLE
RE-ALIGN MACHINE
UTILISATION
XXX
5. Select the RE-ALIGN MACHINE menu using the arrows keys
SET LASER TYPE
TICKLE
RE-ALIGN MACHINE
UTILISATION
XXX
6. Validate
RE-ALIGN X-Y
RE-ALIGN Z
ALIGNMENT MODE
7. Select the RE-ALIGN Z menu using the arrows keys
RE-ALIGN X-Y
RE-ALIGN Z
ALIGNMENT MODE
Autofocus adjustment