20
Accuracy
(cont’d)
AC Voltage
Range
Accuracy* at
20 Hz to 45 Hz
Accuracy* at
45 Hz to 65 Hz
Accuracy* at
65 Hz to 10 kHz
500.00 mV
± (1.2% + 0.4 mV)
± (0.3% + 0.2 mV)
± (0.4% + 0.25 mV)
5.0000 V
± (1.2% + 0.004 V)
± (0.3% + 0.003 V)
± (0.3% + 0.004 V)
50.000 V
± (1.2% + 0.04 V)
± (0.3% + 0.03 V)
± (0.3% + 0.04 V)
500.00 V
Unspecified
± (0.5% + 0.3 V)
± (0.5% + 0.4 V)
1000.0 V
Unspecified
± (0.5% + 3 V)
± (0.8% + 4 V) (65 Hz to 1 kHz)
Range
Accuracy* at
10 kHz to 20 kHz
Accuracy* at
20 kHz to 100 kHz
500.00 mV
± (0.5% + 0.3 mV)
± (2.5% + 0.4 mV)**
5.0000 V
± (0.7% + 0.004 V)
± (3.5% + 0.004 V)**
50.000 V
± (0.7% + 0.04 V)
± (3.5% + 0.04 V)**
500.00 V
± (0.7% + 0.4 V)
Unspecified
1000.0 V
Unspecified
Unspecified
*
From 5% to 10% of range: Accuracy percentage of r 80 digits.
**
From 5% to 10% of range: Accuracy percentage of r 180 digits.
Residual reading less than 50 digits with test leads shorted.
Note: “Digits” refers to the least significant digit (the number in the right-most position).
DC Voltage
Range
Accuracy
Input Impedance
500.00 mV
± (0.02% + 0.02 mV)
10 MΩ, 80 pF nominal
5.0000 V
± (0.02% + 0.0002 V)
10 MΩ, 60 pF nominal
50.000 V
± (0.03% + 0.002 V)
500.00 V
± (0.04% + 0.02 V)
1000.0 V
± (0.15% + 0.2 V)
Summary of Contents for DM-860A
Page 14: ...14 Typical Measurements Voltage Measurement Current Measurement ...
Page 16: ...16 Typical Measurements Temperature ...
Page 26: ...26 ...
Page 39: ...DM 860A 39 Mediciones más comunes Medición de voltaje Medición de corriente ...
Page 41: ...DM 860A 41 Mediciones más comunes Temperatura ...
Page 63: ...DM 860A 63 Mesures types Mesure de tension Mesure d intensité ...
Page 65: ...DM 860A 65 Mesures types Température ...
Page 75: ...DM 860A 75 ...