9 Specification:
Measuring range:
humidity:
0.0
… 100.0 %RH (rec. operating range: 10 – 90 %RH)
temperature:
-25.0
… +60.0 °C
Resolution:
0.1 %RH and 0.1 °C (display and memory)
Accuracy:
(at nominal temperature)
humidity:
≤ ±3 %RH (at range 10 – 90 %RH)
temperature:
± 0.3 °C ± 0.017 * (T - 25°C)
Sensor:
mounted in sensor tube
Display:
LCD display, 10 mm high, 4-digit
Recording interval:
4s to 5h
Measuring value memory:
16.000 values (tempe humidity)
Memory type:
FILLING MEMORY
:
>>
Once the memory is filled with data, the recording will automatically be halted.
RING MEMORY
:
>>
The old data will be overwritten in case of memory overflow.
Recording time:
approx. 18 hours up to 9 years, depending on measuring cycle.
Nominal temperature:
+25 °C
Working temperature:
-25
… +60 °C
Storage temperature:
-30
… +85 °C
Battery:
CR 2032, exchangeable.
Battery service life:
approx. 3 years (if interval is 15 min.), depending on measuring cycle and operating
temperature.
Interface:
Serial interface, 3-pin miniature plug.
Data communication:
via interface converter.
Dimensions / housing:
48.5 x 48.5 x 35.5 mm (H x W x D), plug, fixation flap and sensor tube not included.
Housing made of shock resistant plastic, transparent front made of polycarbonate,
splash water-proof: IP 65 (without protection cap).
Sensor tube:
approx. Ø 15 mm, made of polyamide,
screw-type plastic protection cap for quick response, made of polycarbonate
protection cap: IP40
EMC:
The
T-Logg 160
have been manufactured in accordance with the regulations
concerning EMC (2004/108/EG).
The device meets EN61326-1.
Additional error: < 1%.
Note: avoid ESD in the area of the sensor protection cap!
10 Disposal instructions:
The device must not be disposed in the regular domestic waste. Send the device directly to us (sufficiently
stamped), if it should be disposed. We will dispose the device appropriate and environmentally sound.