Operating Manual EASY

LOG

 80CL 

Page 12 of 14 

12.  Frequently asked questions [FAQ] 

Problem: 

The logger cannot be stopped via keys. 

Possible reason: 

“Logger stop via buttons disabled” has been activated at the configu-
ration of the logger. 

Remedy: 

Deactivate this function by means of GSOFT 

40K (at tab „settings“). 

 

Problem: 

The logger cannot be stopped both via keys and GSOFT 40K. 

Possible reason: 

“Do not stop active logger” has been activated at the configuration of 
the logger. 

Remedy: 

Deactivate this function by means of EASYBus-configurator or 
EbxKonfig. 

Problem: 

The alarm settings or the sea-level correction cannot be 
changed. 

Possible reason: 

The logger has still saved data. 

Remedy: 

Stop logger as recommended by the software (delete data). 

Problem: 

The desired calculated variable cannot be changed. 

Possible reason: 

The logger has still saved data. 

Remedy: 

Stop logger as recommended by the software (delete data). 

Problem: 

A error message is displayed. 

Error message 

Possible reason 

procedure / remedy 

 

measuring range 
exceeded

 

Measuring range of the chan-
nel has been exceeded. 

Operate logger within its specifica-
tions.  

humidity: sensor bedewed 

Twist filter cap carefully off and let 
sensor dry.   

Sensor defective 

Send the logger to the manufac-
turer for repair. 

 

measuring range 
under-run 

Measuring range of this chan-
nel is under-run. 

Operate logger within its specifica-
tions. 

Sensor defective 

Send the logger to the manufac-
turer for repair. 

 

system error 

Recording error 

Reset the error via GSOFT 40K. 
If the error message is displayed 
furthermore send the logger to the 
manufacturer for repair. 

System error 

Send the logger to the manufac-
turer for repair. 

 

sensor error 

Humidity / temperature sensor 
has moved out of its socket  

Check the position of humidity / 
temperature sensor. 

Sensor defective 

Send the logger to the manufac-
turer for repair. 

 

calculation error 

Error message at a variable 
needed for the calculation 
(humidity / temperature / pres-
sure) 

Remedy the reason for error at this 
channel. 

 

 

Summary of Contents for 602773

Page 1: ...electronic GmbH D 93128 Regenstauf Hans Sachs Stra e 26 phone 49 9402 9383 0 fax 49 9402 9383 33 eMail info greisinger de ab f from Version V1 8 Operating Manual EASYLOG 80CL Data logger for humidity...

Page 2: ...logger EASYLOG 80CL is designed for long term monitoring of climate data Not only relative humidity temperature and air pressure can be displayed but also wet bulb tem perature dew point temperature...

Page 3: ...tentials 4 If there is any risk whatsoever involved in running it the device has to be switched off immediately and to be marked accordingly to avoid re starting Operator safety may be at risk if ther...

Page 4: ...play element for max min Logg indicates that logger is recording BAT signals weak battery see chapter 9 AL signals an alarm state at minimum one channel SL signals an active sea level correction for t...

Page 5: ...the display element MAX Note You can immediately switch to the next measuring channel pushing the button again Delete max value Push button for about 5 seconds CLr is displayed and the max values of...

Page 6: ...tart recording Push button for about 5 seconds The main display shows run The secondary display shows no To start recording select YES with the arrow keys button or and verify with Note If no key is p...

Page 7: ...ed if the logger is stopped logger without logger data 7 4 Operating possibilities via software EBxKonfig or EASYBus Configurator The software EBxKonfig V3 9 or EASYBus Configurator V2 0 is needed to...

Page 8: ...ow AL signals an alarm state at minimum one channel The arrow SL signals an active sea level correction 9 2 Logger status messages At stand alone operation no EASYBus connected logger status messages...

Page 9: ...er to the manufacturer for repair 9 4 Alarm and error messages The following messages correspond to a particular measuring channel and are dis played accordingly ALARM LOW The measured value is below...

Page 10: ...at range 10 90 RH Air pressure 1 0 mbar typ 2 5 mbar max Display two 4 digit LCD displays about 7 and 5 8 mm high Interface EASYBus Busload equals to 2 EASYBus elements Battery life time depending on...

Page 11: ...ng transparent screen made of polycarbonate splash proof acc to IP65 excl protection cap Sensor tube about 15 mm made of polyamide Protection cap removable twist plastic head made of polycarbonate IP4...

Page 12: ...oftware delete data Problem A error message is displayed Error message Possible reason procedure remedy measuring range exceeded Measuring range of the chan nel has been exceeded Operate logger within...

Page 13: ...tion diagram for a relay with self induction recuperation diode b Connection information for GR 10 GNR 10 or GNR 232 A GND to be connected to terminal 11 of the GNR NPN to be connected to terminal 10...

Page 14: ...E40 0 01 6C 04 Operating Manual EASYLOG 80CL Page 14 of 14...

Reviews: