E40.0.01.6C-04 

Operating Manual EASY

LOG

 80CL 

Page 7 of 14 

7.3  Operating possibilities via software GSOFT 40K 

You can easily operate the logger via the software GSOFT 40K (> V7.14). This software 
provides a variety of functions to operate the logger. For example: 

•  start and stop the logger 
•  read out the logger data; illustrate them graphically and save them 
•  adjust the alarm boundaries and alarm delay 

*

1

 

•  activate the sea-level correction and enter the actual sea level 

*

1

 

Description: 

The measured barometric air pressure can be converted to the air 
pressure at sea level if the actual height above sea level is entered.   

 

•  activate the function: “logger stop via buttons disabled“ 

*

1

 Please note:  

The alarm settings and the sea-level correction can be only 
changed if the logger is stopped (= logger without logger data).

 

7.4  Operating possibilities via software EBxKonfig or EASYBus-

Configurator 

The software EBxKonfig (> V3.9) or EASYBus-Configurator (> V2.0)

 

is needed to 

change the configuration settings.

  

With that software the following adjustment are additionally possible.  

•  Selection of the unit of the calculated variables

*

2

  

•  Offset and slope adjustment for each measured variable 

*

2

 

•  Activation of the function: “do not stop active logger”

 

*

2

 Please note:  

This can be only changed if the logger is stopped (=logger without 
logger data)

 

8. Battery life time and recording period 

 

Measuring cycle:

 

4 sec 

3 min 

15 min 

5 h 

 

Recording period:

 

11.5 days 

521 days 

7.1 years 

142 years 

 

Battery life time:

 

-- 

-- 

ca. 5 years 

-- 

 
 

 

Please note:

 

Short measuring cycles seriously reduce the battery life time!  
 
In this case you should consider not unplugging the 

EASYBus

-

interface.   
Then the logger is supplied via the interface. 
The internal battery is spared. 

 

 

 

 

Summary of Contents for 602773

Page 1: ...electronic GmbH D 93128 Regenstauf Hans Sachs Stra e 26 phone 49 9402 9383 0 fax 49 9402 9383 33 eMail info greisinger de ab f from Version V1 8 Operating Manual EASYLOG 80CL Data logger for humidity...

Page 2: ...logger EASYLOG 80CL is designed for long term monitoring of climate data Not only relative humidity temperature and air pressure can be displayed but also wet bulb tem perature dew point temperature...

Page 3: ...tentials 4 If there is any risk whatsoever involved in running it the device has to be switched off immediately and to be marked accordingly to avoid re starting Operator safety may be at risk if ther...

Page 4: ...play element for max min Logg indicates that logger is recording BAT signals weak battery see chapter 9 AL signals an alarm state at minimum one channel SL signals an active sea level correction for t...

Page 5: ...the display element MAX Note You can immediately switch to the next measuring channel pushing the button again Delete max value Push button for about 5 seconds CLr is displayed and the max values of...

Page 6: ...tart recording Push button for about 5 seconds The main display shows run The secondary display shows no To start recording select YES with the arrow keys button or and verify with Note If no key is p...

Page 7: ...ed if the logger is stopped logger without logger data 7 4 Operating possibilities via software EBxKonfig or EASYBus Configurator The software EBxKonfig V3 9 or EASYBus Configurator V2 0 is needed to...

Page 8: ...ow AL signals an alarm state at minimum one channel The arrow SL signals an active sea level correction 9 2 Logger status messages At stand alone operation no EASYBus connected logger status messages...

Page 9: ...er to the manufacturer for repair 9 4 Alarm and error messages The following messages correspond to a particular measuring channel and are dis played accordingly ALARM LOW The measured value is below...

Page 10: ...at range 10 90 RH Air pressure 1 0 mbar typ 2 5 mbar max Display two 4 digit LCD displays about 7 and 5 8 mm high Interface EASYBus Busload equals to 2 EASYBus elements Battery life time depending on...

Page 11: ...ng transparent screen made of polycarbonate splash proof acc to IP65 excl protection cap Sensor tube about 15 mm made of polyamide Protection cap removable twist plastic head made of polycarbonate IP4...

Page 12: ...oftware delete data Problem A error message is displayed Error message Possible reason procedure remedy measuring range exceeded Measuring range of the chan nel has been exceeded Operate logger within...

Page 13: ...tion diagram for a relay with self induction recuperation diode b Connection information for GR 10 GNR 10 or GNR 232 A GND to be connected to terminal 11 of the GNR NPN to be connected to terminal 10...

Page 14: ...E40 0 01 6C 04 Operating Manual EASYLOG 80CL Page 14 of 14...

Reviews: