Quick reference guide 

G 1200 series 

Thermocouple    
Quick Response Thermometer 

EN 

Summary of Contents for G 1200 Series

Page 1: ...Quick reference guide G 1200 series Thermocouple Quick Response Thermometer EN ...

Page 2: ... 6 3 The device at a glance 7 3 1 G 1200 series 7 3 2 Display elements 7 3 3 Connections 7 3 4 Operating elements 8 4 Operation 9 4 1 Opening the configuration menu 9 4 2 Adjustment of the measuring input 10 5 Measurement Basics 11 5 1 Sensor device accuracy 11 5 2 Possible measuring errors 11 5 2 1 Immersion depth 11 5 2 2 Surface effects and poor heat transfer 11 5 2 3 Cooling evaporation 11 5 2...

Page 3: ...0 3 18 6 2 Battery 12 6 2 1 Battery indicator 12 6 2 2 Changing battery 12 7 Error and system messages 14 8 Technical data 15 9 Disposal 16 10 Service 17 10 1 Manufacturer 17 10 2 Calibration and adjustment service 17 10 3 Accessories 18 ...

Page 4: ...e disregarding this document disregarding safety notices assignment of inadequately qualified technical personnel and arbitrary modifications of the product This document is entrusted to the recipient for personal use only Any transmission duplication translation into other languages or excerpts from this operating manual require the consent of the manufacturer The manufacturer assumes no liabilit...

Page 5: ... fault free function and operational safety of the product can only be guaranteed if applicable safety precautions and the device specific safety instructions for this docu ment are observed If these notices are disregarded personal injury or death as well as property damage can occur _ DANGER Incorrect area of application In order to prevent erratic behaviour of the product personal injury and pr...

Page 6: ...e is thermometer for type K interchangeable probes It is designed with the appropriate probe for precise and instantaneous temperature measurements in the following media Liquids gases soft plastic materials bulk material Food 2 5 Qualified personnel For commissioning operation and maintenance the relevant personnel must have adequate knowledge of the measuring process and the significance of the ...

Page 7: ...ement of the current temperature or value for min max hold _ Auxiliary display Measurement of the current temperature in min max hold mode with unit 3 3 Connections Thermos couple plug Connection for temperature sensor CAUTION The device ensures extensive protection against splashing water and rain However immersion in water is not possible due to the socket opening Protect the connector opening c...

Page 8: ...min max value Change value of the selected parameter Long press Reset the min max value of the current measure ment Both simultaneously Rotate display overhead display _ Function button Press briefly Freeze measurement Hold Return to measurement display Call up next parameter Long press 2s Start menu configuration ONF appears in the display Operating status device is in measured value display devi...

Page 9: ... off after a selected time in hours minutes during which no buttons have been pressed L TE Backlight OFF Backlight deactivated 0 15 0 30 1 00 2 00 4 00 Automatic shut off of the backlight after a selected time in minutes seconds during which no buttons have been pressed ON No automatic shut off of the backlight UN T Display unit Temperature display in C F Temperature display in F IN T Factory sett...

Page 10: ... the On Off button to switch on the device and open the Adjustment menu 3 The display shows the first parameter Release the down button Parameter Values Meaning _T OF Zero point correction 0 00 No zero point correction 5 0 5 0 Zero point correction in C at F 9 0 9 0 T SL Gradient correction of the temperature 0 00 No gradient correction of the temperature 5 00 5 00 Gradient correction in Formula Z...

Page 11: ...errors can occur due to the heat transmission of the sensor tube if the immer sion depth is too shallow Gas Immerse as far as possible into the gas to be measured so that the measur ing sensor is subjected to a heavy flow 5 2 2 Surface effects and poor heat transfer _ Special measuring sensors are required for this purpose Surface characteristics de sign of the measuring sensor heat transfer and e...

Page 12: ...om the batteries are avoided as a result 6 2 Battery 6 2 1 Battery indicator _ If the empty frame in the battery display blinks the batteries are depleted and must be replaced However the device will still operate for a certain length of time If the BAT display text appears in the main display the battery voltage is no longer adequate for operation of the device The battery is fully depleted 6 2 2...

Page 13: ...should therefore be avoided NOTE Read the following handling instructions before replacing batteries and follow them step by step If disregarded the device could be damaged or the protection from moisture could be diminished 1 Unscrews the Phillips screws A and remove the cover 2 Carefully replace the two Mignon AA batteries B En sure that the polarity is correct It must be possible to in sert the...

Page 14: ...evice Replace battery Send in for repair BAT Battery depleted Battery depleted Replace battery ERR 1 Measuring range exceeded Measurement too high Incorrect measuring probe connected Measuring probe or device defect Stay within allowable measurement range Check measuring probe Send in for repair ERR 2 Measuring range is undercut Measurement too low Incorrect measuring probe connected Measuring pro...

Page 15: ...5 C K Measuring cycle approx 3 measurements per second Display 3 line segment LCD additional symbols illuminated white luminous duration adjustable Display orientation 180 rotatable overhead display Additional functions Min max hold Adjustment Offset and gradient correction Housing Break proof ABS housing Protection rating IP40 sensor socket IP65 IP67 rest of the housing Dimensions L W H 108 54 28...

Page 16: ...cording to Table 1 Additional errors 0 5 FS EN 50581 2012 The device is intended for mobile use and or stationary operation in the scope of the specified operating conditions without further limitations _ _ 9 Disposal Separation by material and recycling of device components and packaging must take place at the time of disposal The valid regional statutory regulations and directives applicable at ...

Page 17: ...the iso calibration certificates These certificates area affordable alternative to the DAkkS calibration certificates and provide information of the traceable reference a list of individual values and documen tation The DAkkS calibration is based on DIN EN ISO 17025 the accreditation basis is recognised worldwide These certificates offer high quality calibration and consistently high quality The D...

Page 18: ...ith silicone handle Ø1 5 mm GF 1TK L3 art no 611299 Compact air probe with silicone handle Ø3 mm GTF 300 art no 600072 Wire probe for measurements in fractions of a second GTF 40K 620 art no 610829 T handle insertion probe for bulk material soil compost Further accessories ST G1000 art no 611373 Protective bag with belt clip GCLIP 1000 art no 475820 Metal belt clip self adhesive G1000_BASE art no ...