7 | Error and system messages
G 1200 series
14 / 18
B-H89.0.0X.DK2-1.0
7
Error and system messages
Display
Meaning
Possible
causes
Remedy
----
No suitable meas-
uring probe con-
nected
Measurement far
outside of the
measuring range
Incorrect measuring
probe
Measuring probe or
device defect
Connect a suitable meas-
uring probe
Measurement leaves the
permissible range
Send in for repair
No display,
unclear char-
acters or no
response when
buttons are
pressed
Battery depleted
System error
Device is defective
Battery depleted
Error in the device
Replace battery
Send in for repair
BAT
Battery depleted
Battery depleted
Replace battery
ERR.1
Measuring range
exceeded
Measurement too
high
Incorrect measuring
probe connected
Measuring probe or
device defect
Stay within allowable
measurement range
Check measuring probe
Send in for repair
ERR.2
Measuring range is
undercut
Measurement too
low
Incorrect measuring
probe connected
Measuring probe or
device defect
Stay within allowable
measurement range
Check measuring probe
Send in for repair
ERR.7
System error
Cold junction out-
side operating range
or defect
Stay within allowable
operating temperature
Send in for repair
SYS ERR
System error
Error in the device
Switch device on/off
Replace batteries
Send in for repair