H24.0.01.6C-07       page 2 of 4

Calibrating of ph-Meter:

Accessories required: calibrating solution for ph7 and ph4 (respectively special accessories ph10, ph12)

How to prepare a calibrating solution:

- fill 2 plastic bottles (special accessories) with 100 ml distilled water (eg battery water) each.

(100 ml will cover the entire cylindrical area, ie approx. 55 mm as of bottle base).

- open ph7 capsule (green) carefully (turn one half of the capsule while pulling and make sure not to spill any solution);

put content (including both capsule parts) into one of the bottles.

- place second capsule for ph4 (orange) respectively ph10 or ph12 including both capsule parts, in second bottle.
The capsule will imbue the liquid with the respective characteristic colour:
orange = ph 4,0; green = ph 7,0, blue = ph 10,0, colourless (white capsule) = ph 12.0
Make sure to prepare buffer solutions in time as they can only be used after 3 hours. Shake well before use.

Calibration of pH-meter:

In order to ensure optimum measuring accuracy, the calibration range should be larger than the measuring range. We
recommend using the following calibration solutions for measurements.

less than pH 7: pH 4,0 and pH  7,0
more than pH 7: pH 7,0 and pH 12,0

Plug in Cinch plug of pH-electrode into corresponding socket and switch on device by means of the lateral slide switch (a
number with decimal point will be displayed on the LCD).
Determine temperature of buffer solution and set accordingly using the central turning knob (controller for temperature
compensation 0 to 90°C): 1 graduation mark = approx.10°C.
Place GPH014 meter on a solid surface; if possible try to avoid holding the meter in your hand during calibration (please refer
to "Points to be observed during measuring").

Setting of the first calibration point:

Carefully remove protection cap from electrode. (Attention: cap contains 3 mol/l KCl).
Rinse electrode with distilled water, dry electrode, place it in buffer solution pH 7.0.
Wait for approx. 20 to 30 seconds (for display to show stable value), then use turning knob at the far right side (pH 7) to set
value to 7.00.

Setting of the second calibration point:

Rinse electrode with distilled water once again, dry electrode, place it in buffer solution pH 4.0.
Wait for another 20 to 30 seconds (for display to show stable value), then use turning knob at the far left side (pH X) to set
value to 4.00.

  Safety Requirements:

This device has been designed and tested in accordance with the safety regulations for electronic devices.
However, its trouble-free operation an reliability cannot be guaranteed unless the standard safety measures and special
safety advises given in this manual will be adhered to when using the device
1. Trouble-free operation and reliability of the device can only be guaranteed if the device is not subjected to any other

climatic conditions than those stated under "Specification".

2. If the device is transported from a cold to a warm environment condensation may cause in a failure of the function. In

such a case make sure the device temperature has adjusted to the ambient temperature before trying a new start-up.

3. If there is a risk whatsoever involved in running it, the device has to be switched off immediately and to be marked

accordingly to avoid re-starting.
Operator safety may be a risk if:
- there is visible damage to the device
- the device is not working as specified
- the device has been stored under unsuitable conditions for a longer time.
In case of doubt, please return device to manufacturer for repair or maintenance.

4.

Attention:

 Do NOT use this product as safety or emergency stopping device, or in any other application where failure of

the product could result in personal injury or material damage.
Failure to comply with these instructions could result in death or serious injury and material damage.

5. The electrodes contain 3 molar KCL (GE103: 1mol/l KNO3), which is acidly.

First-Aid-provisions

After contact with skin:

clean with sufficient water.

After contact with eyes: rinse opened eye with sufficient water, contact oculist
After swallowing:

drink much water. When feeling sick, contact doctor.

Summary of Contents for GPH 014

Page 1: ...egislation for the member countries regarding electromagnetic compatibility 89 336 EWG additional errorr 1 Operating Manual for Digital pH Meter GPH 014 Disposal instruction The device must not be disposed in the unsorted municipal waste Send the device directly to us sufficiently stamped if it should be disposed We will dispose the device appropriate and environmentally sound Points to be observe...

Page 2: ...x 20 to 30 seconds for display to show stable value then use turning knob at the far right side pH 7 to set value to 7 00 Setting of the second calibration point Rinse electrode with distilled water once again dry electrode place it in buffer solution pH 4 0 Wait for another 20 to 30 seconds for display to show stable value then use turning knob at the far left side pH X to set value to 4 00 Safet...

Page 3: ...Use prick lance first Points to be observed during measuring As resistance values resulting from both the pH electrode and from the measuring device are generally very high when conducting pH measurements the electrode should not be hand held but put down during the measuring operation in order to avoid a shift in voltage parameters between the electrode and the measuring device In case there is n...

Page 4: ... if necessary top up with 3 mol l KCl solution through filling hole using a syringe or pipette 2 2 Crystaization of the 3 mol l KCl is not avoidable Crystalized KCl at protection cap or closing collar can easily be removed by fingernail or cloth and is no defect or reason for reclamation 2 3 Dirty electrodes have to be cleaned You will find suitable cleaning agents for the pH glass diaphragm in th...

Reviews: