background image

H24.0.01.6C-07       page 4 of 4

General maintenance and measuring instructions for pH combi-electrodes

(our standard type GE 014)

This pH electrode has been subjected to a series of tests demanding maximum quality standards in all stages of
manufacture. 

Attached electrode can be used for 2 to 12 pH respectively at temperatures between 5 and 45°C and

conductivity > 200 µS/cm.
1.

Make sure to observe the following points to maintain optimum capacity and accuracy of electrode as long
as possible:

1.1. Remove storage-protection cap from pH-glass diaphragm and rinse glass body and pH-glass diaphragm with ordinary

tap water. Then dry glass body and pH-diaphragm with soft tissue.

1.2. Important! Make sure to always keep pH-glass diaphragm in a slightly moist condition. If electrode is not used, the pH-

glass diaphragm has to be immersed into a 3 mol/l KCl solution for storage.
Drying out of the pH-glass diaphragm will affect both its capacity and sensitivity. In order to wet it throughout, put glass
diaphragm in a 3 mol/l KCl solution for 24 hours.

1.3. Please avoid touching of the glass diaphragm as even the slightest damage to, or abrasion of its surface may

negatively affect the capacity of the electrode.

1.4. Make it a rule to always visually check pH-electrodes for any air bubbles that might be enclosed in the pH-glass

diaphragm and the external reference electrode cell. In case of air bubbles being apparent, these can be removed by
shaking the electrode (downwards as for a clinical mercury thermometer)

1.5. In order to avoid a pressure build-up or vacuum in the reference electrode cell, thus ensuring trouble-free penetration,

shift rubber closing collar covering the electrolyte filling hole. For storage, the closing collar has to be properly and
carefully replaced as otherwise the electrolyte will leak.
During measurements the lateral diaphragm should also come into contact with the material to be measured. Minimum
immersion depth for GE 014 is 20 mm, max. 50 mm

1.6. Electrode cable and plug should always be clean and dry as otherwise the electrical insulation may be damaged, this

resulting in incorrect measurements as well as other faults.

2.

Care and maintenance:

2.1. Check liquid level of reference electrolyte at regular intervals; if necessary top up with 3 mol/l KCl solution through filling

hole using a syringe or pipette.

2.2. Crystaization of the 3 mol/l KCl is not avoidable !

Crystalized KCl at protection cap or closing collar can easily be removed by fingernail or cloth and is no defect or
reason for reclamation.

2.3. Dirty electrodes have to be cleaned. You will find suitable cleaning agents for the pH-glass diaphragm in the following

table:
Contamination

Cleaning agent

Various deposits

Light cleaning agent

Inorganic coatings

Commercial liquids for cleaning of glass

Metal compounds

1 mol/l HCI solution

Oil, grease

Special cleaning agents or solvents

Biological coatings containing proteins

1 % pepsin enzyme in 0.1 molar HCI solution

Resin-lignines

Acetones

Highly resistant deposits

Hydrogen superoxide, sodium hypochloride

The distinct materials of the electrode have to be considered when choosing the cleaning agent.
E.g. electrodes with plastics shaft must not be cleaned with solvents. In case of doubt please contact the manufacturer for
further advice.
The same has to be considered when using aggressive or other, non water based agents!

6.

Cheese, fruit, meat

Insertion electrode (

Our type GE 101

).

When conducting measurements in cheese, milk and other products containing proteins, a special cleaning agent has
to be used for cleaning of the electrode (pepsin solution - can be ordered from us).

Normal cleaning: 0.1 molar HCI solution for at least 5 min. or protein cleaning agent.
Generally speaking, the service life of electrodes is 8 to 10 months; with proper care and maintenance their service life
may even be extended to over 2 years. The actual service life is, however, dependant on the individual application
If you can no longer set the pHX-value this could either be an indication that
a)

the electrode is worn out and needs to be replaced, or that

b)

the buffer solution is used up (prepare new solution). Even if treated very carefully during calibration (to avoid carrying
over of buffer solution residuals from one solution to another if electrode is not sufficiently cleaned and dried) buffer
solutions only keep for a certain period of time (approx. 1 month).

Buffer capsules can be kept for an unlimited period of time - we, therefore, recommend to keep a certain number on stock.
PH12-buffer-capsules (white) have to be stored in an exsiccator or stored together with drying agent.
We also recommend to keep a certain amount of 3 mol/l KCl on stock for re-filling.

Summary of Contents for GPH 014

Page 1: ...egislation for the member countries regarding electromagnetic compatibility 89 336 EWG additional errorr 1 Operating Manual for Digital pH Meter GPH 014 Disposal instruction The device must not be disposed in the unsorted municipal waste Send the device directly to us sufficiently stamped if it should be disposed We will dispose the device appropriate and environmentally sound Points to be observe...

Page 2: ...x 20 to 30 seconds for display to show stable value then use turning knob at the far right side pH 7 to set value to 7 00 Setting of the second calibration point Rinse electrode with distilled water once again dry electrode place it in buffer solution pH 4 0 Wait for another 20 to 30 seconds for display to show stable value then use turning knob at the far left side pH X to set value to 4 00 Safet...

Page 3: ...Use prick lance first Points to be observed during measuring As resistance values resulting from both the pH electrode and from the measuring device are generally very high when conducting pH measurements the electrode should not be hand held but put down during the measuring operation in order to avoid a shift in voltage parameters between the electrode and the measuring device In case there is n...

Page 4: ... if necessary top up with 3 mol l KCl solution through filling hole using a syringe or pipette 2 2 Crystaization of the 3 mol l KCl is not avoidable Crystalized KCl at protection cap or closing collar can easily be removed by fingernail or cloth and is no defect or reason for reclamation 2 3 Dirty electrodes have to be cleaned You will find suitable cleaning agents for the pH glass diaphragm in th...

Reviews: