Page 13
DLT 2.0 Differential Level Transmitter
LOE Time
-
Appears when calibrating Sensor B. Press
and
or
to
change the number of seconds without receiving an echo before
the DLT 2.0 displays ECHO LOSS, and Control relays change
state as calibrated under Relay Parameters.
Factory default is 30 seconds and is recommended for most applications,
Minimum is 1 second and maximum is 99 seconds.