Page 35
DLT 2.0 Differential Level Transmitter
DATA LOGGING (Optional)
Setup
Select
Data Logging
from Menu Selections.
Log Site ID
Enter a number from
00
to
99
. The site ID will become
part of the downloaded file name to help distinguish
downloads from different instruments. Press
to store the
setting.
Mode
Select
Level
,
Range
,
Flow
,
HRT
or
Volume
and press
to store the setting. The logger
will auto-configure for 2 or 3 channel logging according
to the selected mode:
Flow
-
3 channels: A Level
B
Flow
Differential
Level
-
3 channels A Level
B
Level
Differential
Range
-
2 channels A Range
B
Range
HRT
&
Vol
- 2 channels A Volume
B
Volume
Set Date
Press
or
to scroll and select Month, Day and Year.
Press
to store the setting.
Set Time
Press
or
to select the current time in Hours, Minutes
and Seconds. Press
to store the setting.
--Data Logging-------
Log Site ID
00
99
Mode
Level
Range
Flow
HRT
Volume
Set Date
Feb 18
/
2008
Mar 19
/
2009
Set Time
11
:
27
:
40
12
:
28
:
41
Interval
10sec
60min
30min
15min
10min
5min
2min
1min
30sec
Log
Start
Stop
Delete
}