Page 36
DLT 2.0 Differential Level Transmitter
Interval
Press
or
to select the logging interval. Press
to
store the setting.
Log
Stop
,
Start
or
Delete
the log file.
Press
or
to
Delete
and
to delete the log file.
Press
or
to
Start
and
to restart the logger.
Note:
You
MUST delete old file and start a new log to apply any changes that
have been made to the
Log Site ID
,
Mode
or
Interval
.
RETRIEVE LOG FILE
Plug a USB Flash Memory Drive (not supplied by Greyline) into the USB output
cable from the instrument. The instrument display will show the USB file
download icon until the log file is transferred to the memory card and then display
file download completed icon. The USB flash drive may be removed.
Download file names will appear in this format:
Tag is set according to the Log Site ID entered in the instrument Data Logging
menu.
Download letter will be A for the first download from an instrument. B for the
second, then C etc. At the letter Z a - character will appear indicating that the
maximum number of downloads for that instrument are on the USB flash drive.
Older files can be erased or moved from the flash memory drive or a new memory
drive can be used.
OPENING LOG FILES
Install Greyline Logger on your PC or laptop. Refer to the Help menu in the
program for detailed instructions.
Select File/Open/Instrument Log (.log) to open the log file from your USB flash
drive.
DLT
_ _
00A
.
LOG
MODEL
TAG DO
W
NLOAD