Page 4
DLT 2.0 Differential Level Transmitter
CONNECTIONS:
POWER INPUT: 100 to 240 VAC 50/60Hz . No adjustments are necessary for voltages within this
range.
Optional DC: 9-32 VDC, 9 Watts. Connect to + and - terminals.
Optional Thermostat and Heater modules are available rated for 115 VAC or 230 VAC.
IMPORTANT NOTE: To comply with CSA/UL electrical safety standards, AC power input and relay
connection wires must have conduit entry to the instrument enclosure. Installation requires a switch,
overcurrent fuse or circuit breaker in the building (in close proximity to the equipment) that is marked
as the disconnect switch.
Risk of electric shock. Loosen cover screw to access connections. Only qualified personnel
should access connections.
Note: Use of instrumentation over 40°C ambient requires special field wiring.
Note: User replaceable fuse 2 Amp 250V (T2AL250V).
!