background image

Page  7

Only if 

Setpoint

 option installed

Setpoint

Act FWD

Setpoint

Res 1.0

°

<MENU> 

Press to advance to next menu item

Use <UP> or <DOWN> to set the Setpoint Action to either FWD (forward) 

or REV (reverse). FWD means that as the setpoint value increases the output 

resistance increases. REV means that as the setpoint value increases the output 

resistance decreases. Factory default is FWD.

Use <UP> or <DOWN> to set the Setpoint Resolution to either 0.5° or 1.0° as 

required. The default is 1.0° which means that the setpoint will increase or 

decrease by 1° for every button press. The display will be 21.0, 22.0, 23.0, etc 

for example. For 0.5 resolution the display will be 21.0, 21.5, 22.0 for example. 

Factory default is 1.0. This applies to both °C and °F operation.

17. SETPOINT ACTION

18. SETPOINT RESOLUTION

<MENU> 

Press to advance to next menu item

Calibrat Ref

1000 ppm

Calibrat

1000 ppm

The calibration reference screen allows user to set the calibration gas level used 

for calibrating the sensor.  Use <UP> and <DOWN> to adjust the ppm value of 

calibration gas.  The minimum value is 400ppm, the maximum value is 2000ppm.  

Increment value is 50ppm for each step.  Press <MENU> to save and advance. If no 

calibration gas is available exposing the sensor to fresh outside can be used and 

calibrate with 400PPM value.  Another method is to determine the CO

2

 ambient 

level using a calibrated unit as a reference and adjust calibration setting to desired 

value

The calibration screen is used for performing calibration of the sensor.  To perform 

calibration of the sensor place calibration hood over sensor housing and apply 

calibration gas.  Wait for several minutes and then press and hold  <UP> or down 

button for five seconds until display shows “Calibrat Done” or “Calibrat Fail”.  Refer 

to Calibration section in this document for more details.

<MENU> 

Press to exit the menu and return to normal mode

<MENU> 

Press to exit the menu and return to normal mode

19. CO

2

 CALIBRATION

20. CO

2

 CALIBRATION

CALIBRATION

Calibration with gas requires a field calibration kit (CDD1-CALKIT-GS) with pressure regulator, tubing, 

sensor adapter and bottle of CO

2

 gas. The certified calibration gas is applied directly to the CO

2

 sensor to 

perform a gas calibration.

Ensure the regulator knob is turned off and then hand tighten the regulator to the CO

2

 gas bottle. A 

length of tubing should be connected between the regulator output and the sensor adapter hood.

Remove the cover of the device to be calibrated to expose the gas sensor. Carefully install the sensor 

adapter hood over the CO

2

 sensor on the PCB. The adapter should fit snuggly and securely over the 

sensor and remain in place during the entire calibration cycle. This will ensure a stable ppm CO

2

 

environment during calibration. Turn the regulator knob on to allow gas to flow to the sensor. The CO

2

 

reading on the LCD will begin to approach ppm level of calibration gas. Wait 3 to 5 minutes until the CO

2

 

reading stabilizes.

Setpoint

Mid 21 

°

C

<MENU> 

Press to advance to next menu item

Use <UP> or <DOWN> to select the Setpoint Midpoint. Units are either C or F 

depending on the Temp Units setting.

Note that the setpoint has + or - 10 steps from the midpoint.

For °C, the midpoint range is 18 to 25°C (resolution = 1°C, default = 21°C) 

For °F, the midpoint range is 64 to 77°F (resolution = 1°F, default = 70°F)

16. SETPOINT MIDPOINT

Copyright © Greystone Energy Systems, Inc.   All Rights Reserved    Phone:  +1 506 853 3057  Web: www.greystoneenergy.com

IN-GE-CD2RMCXXX-04

Summary of Contents for CD2RMC Series

Page 1: ...r conditioned spaces contact factory or other applications Do not exceed device ratings This product is not intended for life safety applications NOTE This CO2 sensor incorporates a Self Calibration feature to correct CO2 sensor drift This feature is recommended for applications where the CO2 will be exposed to fresh air 400 ppm at least one hour per day If the monitored space is occupied 24 hours...

Page 2: ...he enclosure and complete the wiring connection according to the wiring diagram Figure 6 This is a 3 wire sourcing device Connect the plus DC or the AC voltage hot side to the PWR terminal and the common is connected to the COM terminal The device is reverse voltage protected and will not operate if connected backwards It has a half wave power supply so the supply common is the same as the signal ...

Page 3: ...s required If the optional temperature sensor is installed the relay operation may be assigned to operate as a temperature high or low alarm instead of a CO2 alarm All the same relay parameters may be configured to set up the alarm as required SETPOINT MODE The optional setpoint control is operated by two front panel switches for UP and DOWN and the resistive output signal simulates a slidepot act...

Page 4: ... required Factory default is Direct MENU Press to advance to next menu item 5 OUTPUT SIGNAL DIRECTION When either the UP or DOWN keys are pressed the LCD temporarily changes to display the temperature setpoint value in C or F as set by Temp Units in the Setup Menu The display will show the previous temperature setpoint setting such as 22 ºC or 77 ºF If no other action is taken for 5 seconds then t...

Page 5: ...installed Factory default is C Display is used to control what information is displayed on the LCD The selections depend on the installed options Use UP or DOWN to select either CO2 ppm Temp CO2 Temp Rotate or None Factory default is CO2 ppm For a basic CO2 only device the selection is limited to CO2 ppm or None CO2 ppm Only the actual ppm value of the CO2 reading will be displayed Temp Only tempe...

Page 6: ...es in 1 minute increments to prevent false alarms The default is 0 minutes no delay Use UP or DOWN to toggle the relay ON or OFF to test the relay operation Press MENU to reset to OFF and advance or or 12 RELAY SETPOINT 13 RELAY HYSTERESIS 14 RELAY DELAY 15 RELAY TEST RELAY ASSIGNMENT DEFAULT SETPOINT SETPOINT RANGE RESOLUTION Temp High 26 0 C 79 0 F 5 0 to 40 0 C 40 0 to 100 0 F 0 5 Temp Low 15 0...

Page 7: ...several minutes and then press and hold UP or down button for five seconds until display shows Calibrat Done or Calibrat Fail Refer to Calibration section in this document for more details MENU Press to exit the menu and return to normal mode MENU Press to exit the menu and return to normal mode 19 CO2 CALIBRATION 20 CO2 CALIBRATION CALIBRATION Calibration with gas requires a field calibration kit...

Page 8: ...racter OPTIONAL TEMPERATURE SENSOR Type Various Thermistors and RTDs Accuracy Thermistors 0 2 C 0 36 F 25 C 77 F Platinum RTD s 0 3 C 0 54 F 0 C 32 F Nickel RTD s 0 4 C 0 72 F 0 C 32 F Range 0 to 50 C 32 to 122 F Output 2 wire resistive Relay Optional 2 wire output Form A N O 2 Amps 140 Vac 30 Vdc Override Switch Optional 2 wire output 50 mA 12 Vdc max OPTIONAL SETPOINT CONTROL Type Front panel up...

Reviews: