background image

NDD Series - NO2 Transmitter                                                              

Installation Manual 

 

 
Feb 2, 2016

                            

                          NDD-INS-002 

 

  Page 3 

 

Calibration 

The transmitter features a simple snap-mount sensor pcb that is pre-calibrated. This means that the entire sensor pcb can 
simply be replaced with a new calibrated pcb if desired without having to remove the enclosure. This sensor swap requires no 
tools and can be completed in seconds. Simply disconnect the device wiring, remove the old sensor pcb, snap in the new pcb 
and reconnect the device power. There is no need to make any adjustments or apply gas to the transmitter using the sensor 
swap method. 

 

The device may also be calibrated or verified with NO2 gas if required. This requires a field calibration kit consisting of Zero 
and Span of gases, Calibration adapter, Tubing with a calibration cap to cover to the sensor, a miniature screwdriver and a 
digital multi-meter. 
 
Verification can be done without removing the device cover. Simply apply Span gas using the calibration cap attached 
directly to the port on the cover and monitor the output signal. 
 
The device cover must be removed to perform an actual calibration. In this case, the gas calibration cap attaches to the sensor 
fixture inside the enclosure. The sensor must be continuously powered for at least 1/2 hour prior to calibration. Calibrate the 
sensor either in clean air or by applying Zero gas with a flow rate of 0.5 to 1 liters per minute.  Wait 2 minutes and adjust the 
ZERO pot on the sensor board until a 4 mA output is obtained. 
 
Then attach the Span gas supply. Attach the cap to the fixture over the sensor. Slowly turn the valve knob on the regulator to 
let the gas begin flowing. Calibrate the sensor either by applying Span gas with a flow rate of 0.5 to 1 liters per minute.  Wait 
2 minutes and adjust the Span pot on the sensor board until a 20 mA output is obtained. The Span gas need not be the full 
scale concentration, but could be a fraction of this. A half scaled concentration, accordingly, should provide a 12 mA signal. 
 
If the gas cap is too loose on the fixtures, simply place a wrap of electrical tape around the cap to tighten it up. 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
                                    Verify With Cover In Place                                                 Calibrate With Cover Removed 
 
 

Summary of Contents for NDD Series

Page 1: ... measure the NO2 level in the surrounding air and output a proportional value on the 4 20 mA output The NO2 measurement range is 0 to 10 ppm The relay can be used to indicate an alarm condition The trip point for the relay can be programmed to three different settings with an on board jumper If the NO2 level exceeds the trip point then the relay is activated The relay will remain activated until t...

Page 2: ...and does not require a loop power supply This means the signal current is generated by the transmitter and must not be connected to a powered input or device damage will result This is the typical operating mode of a three wire device Check the controller Analog Input to determine the proper connection before applying power Ensure the controller Analog Input AI matches the transmitter output signa...

Page 3: ...d directly to the port on the cover and monitor the output signal The device cover must be removed to perform an actual calibration In this case the gas calibration cap attaches to the sensor fixture inside the enclosure The sensor must be continuously powered for at least 1 2 hour prior to calibration Calibrate the sensor either in clean air or by applying Zero gas with a flow rate of 0 5 to 1 li...

Page 4: ... Pressure Atmospheric 10 Typical Area Coverage 700 m2 7500 ft2 or 15 m 50 ft radius Calibration Adjustments Zero and Span pots Wiring Connections Screw terminal block 14 to 22 AWG Enclosure 2 8 w x 4 1 h x 1 9 d 71 x 104 x 48 mm Power Supply 24 Vdc 20 or 24 Vac 10 non isolated half wave rectified Consumption 50 mA max Output Signal 4 20 mA sourcing Output Drive Capability 550 ohms max 24 Vac dc Re...

Reviews: