background image

CRYO

Therm 

 

 

#4500                                                                          User’s Manual

 

 

                                                                                                                                 Page 6 

 

090205.1.12.16 

Operation: 

 
Operators should have a working knowledge of clinical hydrotherapy and cryotherapy procedures.  In addition, each 
operator should thoroughly read and understand the procedures in this manual before treating patients with 
CRYOTherm. 
 

1. Fill Tank 

The tank must be filled for proper operation of CRYOTherm.  To fill the tank first make sure all control panel switches 
are in the off position and the drain valve is closed.  Press the power switch to on and then slide the fill switches to the 
"Hi" or "Lo" level and allow the water level to rise. Automatic level switches located behind the overflow grille will 
control water levels 

at “Lo” = 4” and “Hi” = 2” to the top of the tank. 

Note: 

These level controls have a ten-second delay to avoid valve chatter. 

Helpful Hint: 

During contrast treatments level switches can be turned off to avoid loss of temperature due to tanks 

being constantly filled with water to replace that discharged from the overflow. 

2. Water Supply Requirements  

CRYOTherm requires connection to a constant water source in order maintain water level and to allow the 
refrigeration system to work properly.  If water supply is lost the electronic unit controller will shut down the 
refrigeration system and will require reset to restore function. 

(see Control Function #13 and the troubleshooting section).

 

 
3. Switch Pump 

“On”  

When the water level covers all jets the pump switch 

(see Control Function #4) 

may be switched 

“on” to begin circulating 

the water.   

WARNING:  Do not run the pump when the tank is empty. 

 

With the dual speed pump switched on water will circulate through the hydrotherapy jets at low speed.  The pump 
may be switched to high speed by depressing the white button on the console top 

(see Control Function #10)

.  After 20 

minutes the pump will automatically resume low speed operation.  The air control valve next to the pump speed button 
can be turned to adjust the amount of aeration during treatment 

(see Control Function #11)

.   

 

Important:

 The pump must be left on at all times for proper operation. Reliable aeration only occurs at high speed. 

 
4. Select Tank Temperature 

To set the temperature of the tank press and hold the 

“Press-To-Set” button 

(see Control Function #8)

, when depressed 

the digital display will switch from reading the actual tank temperature to indicating the set point temperature.  Insert 

the small screwdriver that was supplied into the adjacent 

slot labeled “Adjust 

F” 

(see Control Function #9) 

and adjust 

the set point to the desired temperature.  When the button is released, the digital display will return to reading the 
actual tank temperature.  

5.  Cool Switch 

Slide the Cool Switch to “On” and the CRYOTherm will seek the set point temperature and control at that point.  
Cooling rate is approximately four degrees per hour. 

 
6.  Patient Loading & Unloading 

Users may access the tank via the compartment steps or over the front and side edges around the perimeter of the 
tank. Users also have access to the pump speed button and air control dial during the course of treatment however 
they should never touch the control panel settings beneath the clear plastic cover.  CRYOTherm is designed to 
accommodate all size athletes and support them standing, seated, or kneeling in the tank or sitting anywhere on the 
exterior Corian surface. 

 

Summary of Contents for 4500 CRYOTherm

Page 1: ...e Grimm Scientific Industries Inc Tel 800 223 5395 P O Box 2143 Marietta OH 45750 Fax 740 374 5745 www grimmscientific com e mail grimm ee net Weight Tanks Empty 1 950 Pounds Weight Tanks Full 6 186 Pounds Tank Dimensions 94 L x 48 W x 28 D Tank Capacity 508 Gallons Temperature Range 40 F to 70 F Tank Construction 12 Gauge Type 304 Stainless Steel Exterior Construction Dupont Cameo White Corian Vo...

Page 2: ...nd or users 9 To prevent the transmission of pathogens people with infectious diseases or open wounds should not use the CRYOTherm 10 Thoroughly clean the tanks with a disinfectant at least once a week Refer to the care and maintenance section for further safeguards and instructions for cleaning CRYOTherm use of an antibacterial agent when filling the tanks helps prevent or reduce infection 11 Do ...

Page 3: ...nts alone 17 Thoroughly clean the tanks with a disinfectant at least once a week and particularly after treating open lesions see Maintenance section of this manual Water Temperature Terminology Very Cold 40 F to 55 F 5 C to 13 C Cold 55 F to 60 F 13 C to 16 C Warm 90 F to 98 F 32 C to 37 C Hot 98 F to 104 F 37 C to 40 C Very Hot 104 F to 110 F 40 C to 43 C ...

Page 4: ...isplays water temperature and temperature control setting 8 Press To Set Button Switches digital display from reading the actual water temperature to reading the temperature control setting 9 Adjust Slot Adjusts use small screwdriver supplied temperature control settings when Press To Set button is pressed 10 11 13 12 10 High Speed Button Switches pump speed from low to high for 20 minutes 11 Air ...

Page 5: ...local plumbing codes for specific requirements 4 Cold Water Supply Connection A NPT connection is provided for connection to the cold water supply An internal backflow preventer protects the supply and requires annual inspection This supply fills the tank and cools the condensing unit When the supply pressure fails the high pressure cutout on the refrigeration system will trip making it necessary ...

Page 6: ... is empty With the dual speed pump switched on water will circulate through the hydrotherapy jets at low speed The pump may be switched to high speed by depressing the white button on the console top see Control Function 10 After 20 minutes the pump will automatically resume low speed operation The air control valve next to the pump speed button can be turned to adjust the amount of aeration durin...

Page 7: ...apid Athletes should be instructed to enter the tank slowly allowing the overflow drain to accommodate the displaced water and prevent overflow of water onto the floor and electrical components When exiting CRYOTherm via the compartment steps care should be taken not to splash excessive water onto the surface of the steps or around the control panel The clear plastic cover should remain in place w...

Page 8: ...eel Never use stainless steel cleaner to clean the tanks 3 The solid surface Corian exterior surface of CRYOTherm is also seamless non porous and stain resistant Any stubborn marks that do not easily wipe away can be buffed with a wet Maroon Pad apply light to moderate pressure in a circular pattern 4 Filter cartridges should be cleaned frequently and replaced regularly to assure they remain effec...

Page 9: ...r housing Allow disinfectant to stand on the surfaces for at least 30 seconds before wiping Note longer contact times may be required for certain pathogens refer to product label for specific instructions 6 Examine tank walls exterior surfaces and associated components for damage and visual cleanliness Remove any stains using the techniques described in the Care and Maintenance section of this man...

Page 10: ...ilter housing 4 To clean and disinfect the filter housing spray the inside surface of the housing and lid with pure disinfectant spray or comparable product and allow surfaces to remain wet for at least 30 seconds before wiping and reassembling filter 5 To reassemble insert a new or cleaned filter cartridge into the housing and replace the lid Push the lid down completely to assure that it is full...

Page 11: ... organics come from body oils perspiration proteins wastes cosmetics lotions soaps and other sources that are introduced to CRYOTherm water by users and the surrounding environment Signs of increased organics include foaming cloudy water development of a water line odors and accumulation of debris in the system Source water in certain locales may also exhibit relatively high levels of dissolved me...

Page 12: ...nity and or calcium hardness Etching or staining of surfaces Corrosion Foaming water Eye skin irritation Test kits are also available to evaluate the amount of various mineral and metal contaminants in the source water Your local pool store may also provide this service for you Based on findings the water can be treated to remove the damaging materials If the decision is made to treat the water wi...

Page 13: ...most effectively and reduce the amount you need to use Treat water with ORB 3 Spa Enzyme No Foam approximately 30 minutes after introducing sanitizer NOTE When using chemicals always read the labels and strictly follow the manufacturer s instructions regarding dosage handling storage and disposal All of the products needed to comply with these recommendations can be purchased directly from Grimm S...

Page 14: ... it first using SpaPure Alkalinity Increaser 4 If using a sanitizer add it now and turn the pump s on SpaPure Bromine Tablets in feeder 5 Note Floating dispensers should be removed while the tank is in use 6 30 minutes after introducing the sanitizer treat the water with the appropriate amount of ORB 3 Spa Enzyme No Foam 7 Recheck pH and sanitizer levels periodically throughout the week to assure ...

Page 15: ... rather than repair it This warranty does not cover damage resulting from accidents improper use or maintenance of this product and in no case shall warrantor s liability exceed the original purchase price paid for the product This warranty is expressly in lieu of all other express warranties 3 Repairs Should repairs be needed on your CRYOTherm simply call the factory and ask to speak with a servi...

Reviews: