29
Note
:
Do not use running water to clean the CM100M and remove the accessories if
necessary,an industrial vacuum cleaner and a stable, long-handled brush are best
suited for cleaning the CM100M.
Please cut off the power before cleaning.
6.2 Maintenance
CM100M is maintenance free.
6.3 Functional inspections
Please check the instrument’s grinding chamber and safety switch one time half a year:
First step
:
open the grinding chamber room and get out all grinding setting.
Second step
:
press
“
START
”
key and the instrument can not work any more.
Third step
:
Close grinding chamber room and press”
START
” and instrument work.
Note
:
Please keep contact with
GRINDER
and inform relative distributor if the instrument can
work when the grinding chamber room is opening.
It’s danger to people if the main switch is break down.