32
15.710010-V5.0
Oxiperm 164 150-2500 g en
6.7
Electrical Connection
!
WARNING
The electrical connection must only be carried out by
qualified personnel!
Observe the local safety regulations!
Danger to life! Switch off the power supply before
connecting the mains cable!
Power supply connection
!
WARNING
This must be disconnected from the power supply each
time the system is connected.
A direct connection of the system to the power supply is made using the terminals
1 (L1), 2 (N) and 3 (PE). Depending on the version, the power supply can be 230
V (AC) or 115 V (AC) .
NOTE
Make the electrical connection according to the
enclosed circuit diagram.
CAUTION
Connection to an incorrect power supply may destroy
the system.
➜
Power supply connection: see terminal connection diagram in enclosure
NOTE
When delivered, the terminals 85 / 86 (remote on/off)
and 87 / 88 (alarm - gas warning device ) are shorted.