![background image](http://html1.mh-extra.com/html/grundfos-alldos/oxiperm-c-164-series/oxiperm-c-164-series_operation-and-service-manual_2272215044.webp)
44
15.710010-V5.0
Oxiperm 164 150-2500 g en
NOTE
The entered codes are reset by entering the number
"1998". Then Level 1=0, Level 2=0
●
In the "Basic setting" menu, scroll to page 2 using "Arrow right".
●
Switch to "Code function" menu using "F1" key.
Display:
Code function
1 Level 1
2 Level 2
Switch to operator selection using "F1" key in level 1.
Display:
Level 1 operator
Code: 0
●
Activate input field using "F1" key.
Enter code (max. 4 digits) using the "Arrow right", "Arrow left", "Up" and
"Down" keys.
●
Confirm using "Enter" key.
●
Use the "Enter" or "Escape" key to return to the "Code function" menu.
Switch to "Level 2 Parameters" using "F2" key.
●
Activate input field using "F1" key.
Input of code analogous to level 1.
●
Use the "Enter" or "Escape" key to return to the "Code function" menu.
●
Use "Escape" key again to return to "Basic setting" menu.
7.7.5 Reset Function
●
In the "Basic setting" menu, select the reset function using the "F2" key, or
scroll in the "Basic setting" menu.
Display:
Reset function
Code: xxxx
Input password
●
Switch to the input field using the "F1" key, enter new code using the "Up",
"Down", "Arrow right" and "Arrow left" keys, enter code
Reset code: 6742
●
Confirm using "Enter" key.
●
Use the "Enter" or "Escape" key to return to the "Basic setting" menu.
7.7.6 Date/Time
In the "Basic setting" menu, switch to the "Date/time" menu using the "F3" key.
Display:
Time setting
1 Time
2 Date
3 Summer time
Time
●
Switch to the "Time" menu using the "F1" menu.
Display:
Time setting
New
Current
XX:YY:ZZ
17.07.02
2 Accept