IV. OPTIONAL ACCESSORIES.
IV.1. PHOTOTHERAPY ASSEMBLY.
To turn on the Phototherapy , just press the
I/O
switch.
Electric starters allow an immediate swiching on.
WARNING: C
OVER INFANT
'
S EYES TO PROTECT FROM HIGH LIGHT LEVELS
.
Reset the clock to zero when starting the treatment, to know the exact time for each
patient.
The total time clock will inform the total time that the Phototherapy is being used. In
such way the user will know how long the lamps have been operating.
When the treatment time is over or any time the operator wants to interrupt the
treatment, press
I/O
to switch off the unit.
Phototherapy assembly comprises four DULUX ® L lamps, specially designed for this
kind of treatment.
The electronic starters allow a reliability over 30.000 switching on operations.
The lamps emit at 8 µW/cm
2
/nm, measured at 80 cm distance. For an effective
treatment, 4 µW/cm
2
/nm are required at least.
After operation hours the lamps lose gardually their output power, being recommended
to check periodically the radiation power with the appropriate measuring equipment.
When radiation drops below 4 µW/cm
2
/nm, the lamps should be replaced.
IV.2. RESUSCITATION/ASPIRATION ASSEMBLY.
Provided with peak inspiratory pressure control valve and pressure relief back-up valve,
manometer gauge, patient circuit. Flowmeter with humidifier and Venturi suction
device for aspiration with collecting jar.
Using the NEOPUFF™ Infant Resuscitator
Summary of Contents for NESTOMAT 6050
Page 31: ...GUIDO RAYOS X S A NESTOMAT 6050 31 MAN 012 Sept 00 Ed 1 Rev 3 1 2 2 3 4 5 6 7 8 9 10...
Page 37: ...GUIDO RAYOS X S A NESTOMAT 6050 37 MAN 012 Sept 00 Ed 1 Rev 3...
Page 38: ...GUIDO RAYOS X S A NESTOMAT 6050 38 MAN 012 Sept 00 Ed 1 Rev 3...
Page 47: ...GUIDO RAYOS X S A NESTOMAT 6050 47 MAN 012 Sept 00 Ed 1 Rev 3...