49
AR006089
________________________________________
Measured values:
Temperature and relative humidity measured by
an external Digi/E series probe. The dew-point
temperature
is calculated from the measured
temperature and the relative humidity.
Range, accuracy and response time:
Depending on the connected Digi/E series probe.
Connection mode:
Digi/E series probe. The probe is provided with a 4-pin
M8 ELKA 4008V connector. The length of the Digi/E
probe cable
t
not exceed 15 m.
Display resolution:
Temperature incl. dew-
point temperature: 0.1 °C.
Relative humidity: 0.1 %RH.
Recommended calibration interval:
1 year (according to the connected probe)
AR006090
________________________________________
Measured values:
Internal temperature, relative humidity.
1x External temperature measured by an
external
Guilcor
Pt1000/E probe.
Dew-point temperature calculated from the internal
temperature and relative humidity.
Difference of external temperature and dew point
temperature.
Ranges:
Internal temperature: (-
20 to +60) °C
Relative humidity: (0 to 100) %RH without lasting
condensation
(*2)
.
Dew-point temperature: -
90 to +60 °C
External temperature input:
(-
90 to +260) °C, Pt1000/3850 ppm sensor
Measuring current: approx. 0.5 mA in pulse
length approx. 60 ms
Accuracy
(*1)
:
Internal t
emperature: ± 0.4 °C
Relative humidity:
-
sensor accuracy ±1.8 %RH
(at 23 ºC in the range of 0 to 90 %RH)
-
hysteresis < ±1 %RH
- non-
linearity < ±1 %RH
- temperature error:
0.05 % RH/°C
(0 °C to +60 °C)