50
Dew-
point temperature: ±1
.
5 °C at an ambient
temperature of T< 25
°C and
RH >30 %.
For details see diagrams in Annex 3.
External temperature input (without probe):
± 0.2 °C in the range
-
90 °C to +100 °C
± 0.2
% of the measured value in the range
+100 ° C to +260 ° C
The precision of the device with the connected
temperature probe is determined by the above-
mentioned precision of input and accuracy of
the probe used.
Connection method (external probe):
Two-wire connection with the possibility of
compensation for the cable wire resistance. The probe
is terminated by a 3-pin M8 ELKA 3008V connector.
The method of interconnection is given in Appendix 2.
The recommended length of probes Pt1000/E is up to
15 m, not longer than 30 m. It is recommended to use
shielded cables.
Response time (air streaming velocity approx. 1 m/s)
(*2)
:
Internal temperature: t63 < 2 min, t90 < 8 min
(temperature jump 20 °C)
Relative humidity: t63 < 45 s, t90 < 4 min, t99 <4 h.
(humidity jump 30 %RH, constant
temperature)
Display resolution:
Internal temperature incl. dew-
point temperature: 0.1 °C
Relative humidity: 0.1 %RH
External temp. input: 0.1 °C, 16
-bit converter range
Recommended calibration interval:
1 year
(*1)
During battery charging temporary measurement
accuracy deterioration may be found.
(*2)
Important information is provided in the chapter
Manufactured models
–
AR006090
AR003954
________________________________________
Measured values:
Internal temperature, relative humidity, barometric
pressure and the concentration of CO
2
in the air. Dew-
point temperature calculated from the internal
temperature and relative humidity.