13
3. Control of the electrical state of the product
• Check electrical functions and signals.
• Check the wiring for faults and defects.
• Check the cable bushings.
• Check connection options, connectors, etc.
• Check the seal on the control box.
4. Control of the mechanical state of the product
• Clean dirt and other impurities from the product.
• Inspect and assess vital parts of the product.
• Replace any parts of the product which are defective or worn.
• Lubricate the product.
5. Go through point 2 again, to check that everything works
6. Did any new problems arise/were new problems noted in point 5?
• If new problems arose, go back to point 2.
• If no new problems were noted, conclude the inspection.
5 .04
Trouble Shooting
1. Is the plug from the mains network in the socket?
• Put the plug in the socket.
2. Is the socket switched on?
• Switch the socket on.
3. Is there power in the socket, the fuse may have blown?
• Check the fuses and change the fuse if necessary.
4. Is the emergency stop activated?
• The emergency stop is released by turning the switch until it springs up.
5. Can the lifting platform mounted with safety nosing list go up but not down?
• Clean any impurities from the safety nosing strip.
• Check the safety nosing strip for other blockages.
6. Are the connecting plugs between the source of power and the lifting platform
connected?
• Connect the plugs.
7. Is the platform overloaded?
• Remove some of the load from the lifting platform.
• Observe the maximum load indication of the platform.
If the above points do not solve the problem, please contact the Guldmann
Service Team.