5.5 MISC Menu
APS-1102 Programmable AC/DC Power Source
5-29
Table 5-21. PROGRAM Screen Items (2/3)
Setting Name
Item
Operation
FREQ
AC frequency
Sets the AC frequency.
Setting range:
1.0 Hz to 550.0 Hz
Setting resolution: 0.1 Hz
AC frequency
operation type
Sets the method used to change to the specified frequency.
Select one of the following three options.
CONST: Output specified value
KEEP: Output value at end of previous step
SWEEP: Start by outputting value at end of previous
step, then perform a linear sweep to output then
specified value at the end of the current step
WAVE
Waveform
Selects arbitrary waveform to be output.
Setting range: SIN/SQU/ARB1 to ARB16
Waveform operation
type
Sets the method used to change to the specified waveform.
Select one of the following two options.
CONST: Output specified waveform
KEEP: Output previous step waveform
PHS
Phase
Sets output initial phase.
Setting range:
0.0
to 359.9
Setting resolution: 0.1
Phase operation type
Sets the method used to change to the specified phase.
Select one of the following two options.
CONST: Output specified phase
KEEP: Output previous step phase
SYNC CODE
Step synchronized
output
This sets the code (2 bits) to be output in sync with step
execution. The sequence step transitions can checked by
monitoring the step synchronized output.
Bit 0 is output from pin 5 (DOUT5) of the external control
I/O connector and bit 1 is output from pin 6 (DOUT6).
Select one of the following four types of step synchronized
output (the numbers are displayed in the order “bit 1, bit
0”).
During Idle mode, these values are “0, 0”.
0, 0
0, 1
1, 0
1, 1
Step synchronized
output operation type
Specifies the output method for step synchronized output.
Select one of the following two options.
CONST: Output specified code
KEEP: Output same code as at previous step
STOP PHS
Step end phase
enable/disable
Specifies step end phase as enabled or disabled. Select one
of the following two options.
DISABLE: Disable (phase depends on step time)
ENABLE: Enable
Step end phase
Sets step end phase.
Setting range:
0.0
to 359.9
Setting resolution: 0.1