6.3
Command Descriptions
APS-1102 Programmable AC/DC Power Source
6-25
MEASure[:SCALar]:CURRent:AMPLitude:MAX?
This returns the output current peak value (hold) in Apk units. The held output current peak value
full scale is 45.0 Apk. If the measured value exceeds the full scale, 99.9 Arms is returned as a
measurement error.
Syntax:
MEASure[:SCALar]:CURRent:AMPLitude:MAX?
Parameters:
None
Query
example:
MEAS:CURR:AMPL:MAX?
Response format:
<NR2>
Response
example: 26.5
MEASure[:SCALar]:CURRent:AMPLitude:RESet
This resets the stored output current peak value (hold) to 0 Apk.
Syntax:
MEASure[:SCALar]:CURRent:AMPLitude:RESet <NR1>
Parameters:
1
Command example:
MEAS:CURR:AMPL:RES 1
Response format:
None
Response example:
None
MEASure[:SCALar]:CURRent:AVErage?
This returns the output current average value (AC+DC average value) in A units. The output
current average value measurement full scale is ±15.00 A. If the measured value exceeds the full
scale, 99.99 A is returned as a measurement error.
Syntax:
MEASure[:SCALar]:CURRent:AVErage?
Parameters:
None
Query
example:
MEAS:CURR:AMPL?
Response format:
<NR2>
Response example:
1.2
MEASure[:SCALar]:CURRent:CREStfactor?
This returns the output current crest factor. The crest factor measurement full scale is 50.00. If the
measured value exceeds the full scale, 99.99 is returned as a measurement error.
The crest factor is calculated as “current peak value/current RMS value”.
Syntax:
MEASure[:SCALar]:CURRent:CREStfactor?
Parameters:
None
Query example:
MEAS:CURR:CRES?
Response format:
<NR2>
Response
example: 1.41