9.6
Sequence Function
APS-1102 Programmable AC/DC Power Source
9-10
9.6
Sequence Function
Output parameters can be changed abruptly or swept in sequence.
Signal source is internal signal source only.
Sequence function settings are stored in memory with battery backup.
Number of sequences
1 sequence per operation mode (AC or AC+DC mode) and output range (100 V or 200 V)
Number of steps
1 to 255 (within 1 sequence)
Step time
0.1 ms to 999.9999 s (resolution 0.1 ms)
Operations within step
Constant, keep, or linear sweep
Parameters
DC voltage, AC voltage, frequency, waveform, and step synchronized output (2 bits)
Jump times
1 to 999 or continuous
Sequence control
Start:
Starts a sequence.
Stop:
Stops a sequence.
Hold:
Maintains current step output. Restarted when a sequence is started.
Branch:
Branches to the specified step.