2.4
Simple Operation Checks
APS-1102 Programmable AC/DC Power Source
2-9
2.
Panel settings
Set the APS-1102 as shown below.
See
“3.4 Main Operation Example”
, for descriptors
of the settings methods.
Table 2-1. Panel Settings for Operation Check
Item Setting
Output mode
AC-INT mode
Output voltage range
100 V
AC output voltage
100 Vrms
Selected measurement display RMS
3.
Press
the
OUTPUT
key to turn on output.
OUTPUT
OUTPUT
Output is off.
Output is on.
⇒
Check that the control panel MEASURE screen shows 100 Vrms as the AC output voltage.
Figure 2-3. Operation Check
⇒
If the normal values are not displayed, it may be due to an operation fault. Contact Good Will
or a Good Will representative.
4.
Turn off the power, check that OUTPUT off is displayed, then press the power switch to the bottom
(
) position. This shuts off the power supply and turns off the power source unit.
Before turning off the power, make sure that the output on/off LED is off.
Output on/off LED is on.
Check display.
!
CAUTION