4.4
Using the Sequence Function
APS-1102 Programmable AC/DC Power Source
4-27
f) Step synchronized output
The code that is specified in the step being executed is output to the external control I/O connector.
However, “0” is output during a normal output step.
Step 1
Step 2
Step 3
Step synchronized output
Time
Step 4
Step 5
0
1
[Step synchronized output]
0
1
0
0
1
Bit 0
Figure 4-10. Example of Step Synchronized Output
g) Step end phase setting
When the step end phase is set as enabled, the step execution time is as described below.
The step execution time = Set step time + Time for superimposed AC phase to become 0
After the set step time elapses, the setting for that step is retained in the subsequent operation.
Figure 4-11. Sequence Operation during Wait for Phase 0°
Value set for this step
Set value is
retained.
SWEEP
Time set by [step time]
Wait time for
0
phase