APS-7000 Series User Manual
64
5.
Press the
Range
key repeatedly to set the
voltage range for the Vset parameter. The range
will be shown in the top corner, which indicates
that the test will be executed within this voltage
range.
Range
Range LO(250V), MD(500V), HI(1000V) , Auto
6.
Go to the
Vset
setting and set the Vrms level of
the step. If you input a Vset value that is not
within the voltage range, the input value will
be ignored.
Not applicable for Trans1 and Trans2.
Vset
0.00 ~ 1000Vrms (range dependent), Auto
7.
Go to the
Fset
setting set the frequency of the
step. Not applicable for Trans1 and Trans2.
Fset
45.00 ~ 500.0Hz
8.
Lastly, go to the
Repeat
parameter to select the
number of times the test will repeat the
Normal1-Trans1-Abnormal-Trans2-Normal2
sequence of steps. A value of 0 will set the
number of repetitions to infinite.
Repeat 1 ~ 9999, 0(infinite)