GDM-397/GDM-461 User Manual
19
GDM-461:
Displays the true rms value.
5.
Press
Hz% (GDM-397)
or the
yellow button
(GDM-461)
to obtain the
frequency and duty cycle value.
Input Amplitude: (DC offset is zero)
Input Amplitude:≥range×30%
Frequency respouse:GDM-397: ≤400Hz, GDM-461: ≤ 1kHz
Note
If the value of the current to be measured is unknown, set to the highest range
and reduce the range step by step until a satisfactory reading is obtained.
For safety, the measurement time for currents greater than 5A (>5A) should be
less than 10 seconds and the interval time between 2 measurements should be
greater than 15 minutes.
When current measurement has been completed, disconnect the connection
between the testing leads and the circuit under test, and remove the testing leads
from the input terminals of the Meter.