GDM-397/GDM-461 User Manual
21
Warning
To avoid damaging the Meter or the DUT, disconnect power from the circuit and
discharge all high-voltage capacitors before measuring resistance.
To avoid harm to yourself, do not input voltages greater than DC 60V or AC 30V.
To measure resistance, connect the Meter as follows:
1.
Insert the red test lead into the
Ω
terminal and the black test lead into the
COM
terminal.
2.
Set the rotary switch to
Ω
;
resistance measurement (
Ω
) is the default or press
the
BLUE
button to select
Ω
measurement mode.
3.
Connect the test leads to the DUT. If there is a lead on the resistor or SMT
resistor, it is more convenient to use the included multi-purpose socket to carry
out testing.
4.
The measured value shows on the display.