OPERATION
33
Temperature Measurement (GDM-533)
Wire Connection
Diagram
Steps of
Measurement
1.
Turn the function dial to the °C°F position.
2.
Insert the K-type thermocouple into the
adapter socket, and then insert the adapter
socket into the input terminals.
3.
Bring the temperature sensing end of the
thermocouple close to the object surface under
test.
4.
Read the Celsius temperature value on the
display after it gets steady.
5.
Short press the SELECT button to switch
between °C and °F.