SPECIFICATIONS
41
S
PECIFICATIONS
General Specifications
1
Max voltage between the input terminal and COM terminal: Please refer to
the description of input protection voltage for each range.
2
mA/
μ
A input terminal protection: (CE) 600mA 1000V fast-acting fuse,
Φ
6x32mm
3
A input terminal protection: (CE) 11A 1000V fast-acting fuse,
Φ
10x38mm
4
Max display: 22000 (
GDM-541
), 6000 (
GDM-533
)
5
Analog bar: 46 segments (
GDM-541
) (conversion rate: 30 times/s), 31
segments (
GDM-533
)
6
Refresh rate: 2~3 times/s
7
Range: Auto/Manual
8
Polarity display: Auto
9
Over-range Indication: OL
10
Low battery indication: “ ” is displayed.
11
Operating temperature: 0°C~40°C (32°F~104°F)
12
Storage temperature: -10°C~50°C (14°F ~122°F)
13
Relative humidity:
≤
75% at 0˚C~30˚C;
≤
50% at 30˚C~40˚C
14
Operating altitude:
≤
2000m
15
Electromagnetic compatibility: Conforms to EN61326-1:2006 and
EN61326-2-2:2006 standards
16
Battery: 4×1.5V AAA
17
Dimensions: 186mm×89mm×49mm
18
Weight: 420g (GDM-541), 400g (GDM-533) (with battery)
19
Safety standard: IEC 61010-1: CAT III 1000V/CAT IV 600V
20
Certification: CE
21
Pollution degree: 2
22
Information of usage: Indoor and outdoor