GDM-541/533 User Manual
44
AC Voltage
GDM-541
Range
Resolution
Frequency Response
Accuracy
220.00mV 0.01mV
40Hz~1kHz
± (1.0%+10)
1kHz~10kHz
± (1.5%+30)
2.2000V
0.1mV
40Hz~1kHz
± (0.8%+10)
1kHz~10kHz
± (1.2%+50)
40Hz~100Hz (LPF)
± (1.2%+50)
22.000V
1mV
40Hz~1kHz
± (0.8%+10)
1kHz~10kHz
± (1.2%+50)
40Hz~100Hz (LPF)
± (1.8%+50)
220.00V
10mV
40Hz~1kHz
± (0.8%+10)
1kHz~10kHz
± (2.0%+50)
40Hz~100Hz (LPF)
± (2.0%+50)
1000.0V
0.1V
40Hz~1kHz
± (1.2%+10)
1kHz~10kHz
± (3.0%+50)
40Hz~100Hz (LPF)
GDM-533
Range
Resolution Frequency Response Accuracy
60.00mV
0.01mV
40Hz~1kHz
± (1.2%+5)
600.0mV
0.1mV
40Hz~1kHz
± (1.2%+5)
6.000V
0.001V
40Hz~1kHz
± (1.0%+3)
60.00V
0.01V
40Hz~1kHz
± (1.0%+3)
± (1.0%+3)
600.0V
0.1V
40Hz~1kHz
1000V
1V
40Hz~1kHz
± (1.2%+5)
LoZ ACV 600.0V 0.1V
40Hz~1kHz
± (2.0%+5)
LoZ ACV 1000V
1V
40Hz~1kHz
± (2.0%+5)