SPECIFICATIONS
49
Capacitance
GDM-541
Range
Resolution
Accuracy
22.000nF
1pF
± (3.0%+5)
220.00nF
10pF
2.2000μF
100pF
22.000μF
1nF
220.00μF
10nF
± (4.0%+5)
2.2000mF
100nF
22.000mF
1μF
± (10%+5)
220.00mF
10μF
± (20%+5)
GDM-533
Range
Resolution
Accuracy
60.00nF
10pF
± (3%+5)
600.0nF
100pF
6.000μF
1nF
60.00μF
10nF
600.0μF
100nF
6.000mF
1μF
± (10%+5)
60.00mF
10μF
Overload protection: 1000V
Measurement result = displayed value
–
capacitance of open-
circuit test leads
For capacitance ≤1μF (
GDM-533
) and ≤22nF (
GDM-541
), it is
recommended to use the REL mode to deduct the open circuit
reading.
Accuracy guarantee: 1%~100% of range
For range
s of 2.2μF and below, when the accuracy is ≤3%, 10
digits should be added (
GDM-541
).
For ranges of 60mF (
GDM-533
) and 220mF (
GDM-541
), the
measurement time is about 20s.